Aseptico ADU-04ST Operation And Maintenance Instruction Manual Download Page 6

6.

UNIT OPERATION: 

Handpieces - 

Connect the proper high-speed and low-speed 

handpieces to their appropriate connector.  Switch the handpiece 

selector tog-gle to the desired handpiece. Depress foot control 

for operation. 

3-Way Air/Water Syringe -

 Depress the right button for air 

operation, and the left button for water operation. Depressing 

both buttons will create a mist. The syringe features quick-

change autoclavable tips: To remove a tip, press on the locking 

collar surrounding the tip socket and pull the used tip straight out 

of the socket (see Fig 3). To insert a new tip, press locking collar 

and push tip into socket as far as it will go. Release collar and 

gently tug on tip before using to ensure that tip is securely locked 

into socket.

Optional Saliva Ejector 

- Remove the saliva ejector valve from 

its holder and insert a standard saliva ejector tip. Turn the low 

vacuum toggle switch ON. Waste from the vacuum system will 

collect in the attached plastic bottle. When the waste container is 

3/4 full, it must be emptied to prevent a back-up in the system.

Optional HVE System

 - Remove the vacuum hose from its 

holder and insert a standard oral evacuator tip. Depress the 

button switch on the side of the HVE valve for vacuum. Waste 

from the vacuum system will collect in the attached plastic bottle. 

When the waste container is 3/4 full, it should be emptied.

DEPRESS LOCKING 

COLLAR

FIG. 3

CLEANING:

Syringe Tip Sterilization:

1.  Remove contaminated syringe tip.

2.  Remove all visible signs of contamination before 

autoclaving.

3.  Autoclave tip at 132° C (270° F) for ten minutes.

4.  Sterilize between each patient use.

NOTE:

 Since only the tips can be autoclaved, it is recommended 

that the Air/Water Syringe be bagged with a disposable, single-

use plastic sleeve between each patient use.

Saliva Ejector

 - The valve assembly on the saliva ejector snaps 

apart at the swivel. After each use, the vacuum system should 

be cleaned by running clean water or vacuum system cleanser 

through the evacuator hose.

HVE System

 - After each use, the high volume evacuator 

system should be cleaned by running clean water or a vacuum 

system cleanser through the evacuator hose.

Vacuum System Waste Bottles

 - Waste from the vacuum 

systems will collect in their respective plastic waste bottles. 

When the waste containers are 3/4 full, they should be emptied. 

After each use, the vacuum systems should be cleaned by 

running clean water or a vacuum system cleanser through the 

evacuator hoses. 

CAUTION:

 Use only NON-foaming cleansers 

in the vaccum lines.Thoroughly clean the waste bottles.

Water Lines -

 Disinfect the water lines weekly. Prepare a 1:10 

bleach solution (1 part household bleach to 10 parts water). 

Remove water reservoir and discard residual water. Replace 

empty water supply tank and air purge all waterlines. Fill water 

supply tank with bleach solution. Run bleach solution through 

all lines. Allow bleach solution to stand in lines for 10 minutes. 

Remove water supply tank and discard bleach. Flush water 

supply tank and all lines thoroughly with clean water. Air purge 

and leave lines dry until next clinical use. 

Summary of Contents for ADU-04ST

Page 1: ...ADU 04ST Manual Dental Control System OPERATION and MAINTENANCE INSTRUCTION MANUAL HVE and Saliva Ejector options...

Page 2: ...n 6 Figure 5 Air Water Syringe 6 Cleaning 6 Warranty 7 SPECIFICATIONS Case Size 11 in x 11 in x 40 in Case Weight 18 lbs Power Source 80 100 PSI Compressed Air To prevent injury to people and damage t...

Page 3: ...trol NWS 8 Self Contained Clean Water System AA 61 Water Connector for scalers 25 Foot Extension Line OPTIONS AA 21ST High Volume Evacuator HVE AA 25ST Saliva Ejector System Fiber Optics Kit for fiber...

Page 4: ...ttach saliva ejector hose to the quick connect located on top of waste container and then place ejector valve into the holder located on left side of unit To remove waste container assembly push in on...

Page 5: ...ir Water Syringe and high speed handpiece Saliva Ejector On Off Toggle Optional Turns vacuum On Off to the saliva ejector valve AA 21ST High Volume Evacuator Optional Provides vacuum flow to the High...

Page 6: ...ted syringe tip 2 Remove all visible signs of contamination before autoclaving 3 Autoclave tip at 132 C 270 F for ten minutes 4 Sterilize between each patient use NOTE Since only the tips can be autoc...

Page 7: ...s to notify Aseptico s Customer Service department promptly Customer Service will provide Return Material Authorization RMA instructions usually directing that the product be returned for service ship...

Page 8: ...8 P O Box 1548 Woodinville WA 98072 8333 216th Street SE Woodinville WA 98072 425 487 3157 800 426 5913 www aseptico com info aseptico com PRINTED IN THE U S A P N 420261 Rev O ECO 15498 04 2022...

Reviews: