Operating Manual
GageMux
®
www.spcanywhere.com
11
(248) 370-9919
DIPSW1
•
Switches 1, 2, 3, and 4—Settings for Flow Mode, Foot Switch,
Output String Formats, and Baud Rate
DIPSW1
Description
Switch
1
Switch
2
Switch
3
Switch
4
Flow mode, Full
comma delimited
format, 9600 baud,
Switch 1 overrides
2, 3, and 4
Closed
Open or
Closed
Open or
Closed
Open or
Closed
1
2
3
4
OPEN
Normal foot switch
mode, Full comma
delimited format,
9600 Baud
Open
Open
Open
Open
1
2
3
4
OPEN
Normal foot switch
mode, Reading only
format, 9600 Baud
Open
Open
Open
Closed
1
2
3
4
OPEN
Normal foot switch
mode, Mitutoyo
Mux-10 format,
9600 Baud
Open
Open
Closed
Open
1
2
3
4
OPEN
Global foot switch
mode, Full comma
delimited format,
9600 Baud
Open
Open
Closed
Closed
1
2
3
4
OPEN
Global foot switch
mode, Reading only
format, 9600 Baud
Open
Closed
Open
Open
1
2
3
4
OPEN
Global foot switch
mode, Mitutoyo
Mux-10 format,
9600 Baud
Open
Closed
Open
Closed
1
2
3
4
OPEN
Foot switch
triggered mode
Open
Closed
Closed
Open
1
2
3
4
OPEN
Normal foot switch
mode, Reading only
format, 4800 Baud
Open
Closed
Closed
Closed
1
2
3
4
OPEN
Operating Manual
GageMux
®
www.spcanywhere.com
12
(248) 370-9919
DIPSW2
•
Switches 1, 2, 3, 4—Normal or MAX/MIN/TIR Mode
DIPSW2
Description
Switch
1
Switch
2
Switch
3
Switch
4
Port 1
Normal Mode
Open
---
---
---
1
2
3
4
OPEN
Port 1
Max/Min/Tir
Closed
---
---
---
1
2
3
4
OPEN
Port 2
Normal Mode
---
Open
---
---
1
2
3
4
OPEN
Port 2
Max/Min/Tir
---
Closed
---
---
1
2
3
4
OPEN
Port 3
Normal Mode
---
---
Open
---
1
2
3
4
OPEN
Port 3
Max/Min/Tir
---
---
Closed
---
1
2
3
4
OPEN
Port 4
Normal Mode
---
---
---
Open
1
2
3
4
OPEN
Port 4
Max/Min/Tir
---
---
---
Closed
1
2
3
4
OPEN