Figure 2. PCB with jumper to configure OUT2 for current 

output 4─20mA or voltage output 2─10VDC

 

 

Figure 3. Enlarged picture of the PCB with the jumper set to current output (left position)

 

 

 

Figure 4. Enlarged picture of the PCB with the jumper set to voltage output (right position)

 

 
 

Self-diagnostics 

 
The  system  contains  a  complete  self-diagnostic  procedure  that  is  executed  automatically  when  the 
sensor is in operation. The yellow LED is lit if an error is found. Sensors with display show a 

wrench

 if 

an error is found. The wrench is shown and the  yellow LED is lit during the first seconds after power 
up and if the measuring range exceeded. They are automatically turned off when the sensor returns to 
normal operation. The output OUT2 indicates the same information by setting the output voltage to 1V 
or 2 mA.

 

 

Maintenance 

 
The

 

eSENSE  is  basically  maintenance  free  in  normal  environments  thanks  to  the  built-in  self-

correcting 

ABC 

algorithm.  Discuss  your  application  with  Senseair  in  order  to  get  advice  for  a  proper 

calibration strategy. 

 

 