4.3 Voltage output
5. Work interface and interface operation
July 2022
The voltage output of the AFM0725 is a linear output. The relationship between voltage and flow is: Voltage (V)
= 1 + Instantaneous flow (L/min) × 0.16
Figure 8. The cumulative flow
interface menu page has 4 levels, and the function description of each level is shown in Table 9.
When the lock screen icon (ÿ in Figure 7) is on, press and hold the setting key for 5 seconds to unlock and enter the menu page, and press the up and down keys to select the parameters to
be set. If there is no operation within 30 seconds, the system enters the lock screen state. When the alarm switch status icon (ÿ in Figure 7) is on, it means the alarm function is turned on, and when
there is no icon, it means the alarm is turned off. To view the cumulative flow, on the main page, press the down key to jump to the cumulative flow display interface, as shown in Figure 8.
Figure 7. Introduction to screen icons
ÿ is the 485-communication address; ÿ is the fluid type; ÿ is the alarm switch status; ÿ is the 485 communication
status; ÿ is the lock screen icon; ÿ is the instantaneous flow; ÿ is the instantaneous flow unit.
10 / 12