![Asahi KASEI Senseair LP8 Handling Manual Download Page 2](http://html1.mh-extra.com/html/asahi-kasei/senseair-lp8/senseair-lp8_handling-manual_2980214002.webp)
Document Rev
Page
ANO1332
4
2 (2)
≥1.0
≥1.0
≥1.5
[mm]
Storage
Storage in sealed ESD bags.
Storage temperature: -40 – 70°C
Requirements on storage environment: in normal IAQ environments corrosive environments are excluded.
Inspection - verification
Transport, handling or assembly may affect calibration.
A host system counts ABC (Automatic Baseline Correction) period itself and has to write ABC command to
the “Calculation Control byte” when ABC period (8 days) expires. Contact Senseair for further information!
Inspect and perform zero calibration after transport.
NOTE:
≥1.0mm minimum distance
from OBA to cover/PCB.
NOTE:
No gluing or moulding on OBA.
NOTE:
≥1.0mm minimum distance
from OBA to cover/PCB.
[mm]
[mm]
NOTE:
≥1.5mm minimum distance
from particle filter to cover/PCB.