CST. CO., LTD.
18
Abduction / Adduction
5.1.3
How to adjust device for Flexion/Extension of Elbow
<Fig 5.1.6>
Step 1. A patient is sitting comfortably in the chair and has his back tightly
attached to the chair back.
Step 2. Pull out the fixture of arm rotator (1) and adjust the location of M1 to the
patient’s shoulder. And then release the fixture of arm rotator to fix it.
Step 3. Loosen the shoulder height adjusting screw (2), and adjust the height of
the axis of M1 to fit to
the patient’s shoulder. And then tighten the screw
to fix it.
Step 4. Loosen the shoulder width adjusting screw (3), and adjust the width of
M1 to fit to
the patient’s shoulder. And then tighten the screw to fix it.
Step 5. Use hand controller to adjust the angle of M1 to degree 45
°
or less.
Step 6. Adjust the length of hand holder (5).
Setting of Elbow
Flexion / Extension
Caution : Before changing to another exercise, set the angle of M2 to 0 °.