1.  Clean with soap & water or a mild cleaning

agent after every use (or at least once a week)

2.  If parts are contaminated by pathogenic or biologic

material we recommend:

– For disinfection: Helipur

®

 H plus N from

B. Braun Melsungen

3.  Dry immediately after cleaning

4.  Lightly grease aluminium parts after drying using

acid-free grease, e.g. vaseline

5. With aerosol-sealed rotors and bio safety systems:

– Clean & check sealing rings regularly (weekly)

– Replace sealing ring immediately upon indication of

crack formation, embrittlement or abrasive wear

– Lightly grease seal ring with acid-free grease,

e.g. vaseline

6. Rotor and Motor Shaft (clean once a month):

– Disassemble the rotor and clean

– Lightly grease the motor shaft

7.  Checked all parts on a monthly basis for

corrosion damage

RECOMMENDED CLEANING PROCEDURE

Hettich Centrifuge –

 Rotors and Accessories

© Arthrex GmbH

, 201

6

.

 All rights reserved. 

LI2-00016-EN_A

Rotors and attachments may no longer be used upon indication of wear and tear or corrosion!