Centrifuge Preventative Maintenance Schedule

Taken from the operator manuals supplied with each centrifuge

        Pull the mains plug before cleaning

Using detergents follow the instructions of the Hettich Operating Instructions.

Centrifuge Bowl Seal

Cleaning

regularly

Rubber Maintenance

after cleaning

Maintains long life for seal.

Always inspect seal for damage.

Rotor

Cleaning

at least once a week 

Lube

after cleaning

Secured

weekly

Maintains good lab practices for cleanliness. Helps 

maintain good results. Check rotor fixing screw for 

safety. Always inspect rotor for damage.

Buckets

Cleaning

at least once a week 

Lube

after cleaning

Maintains good lab practices for cleanliness. Helps 

maintain good results. Always inspect buckets for 

damage.

Motor Shaft

Cleaning

at least once a month 

Lube

at least once a month 

Helps to ensure that the rotor can be removed  

for maintenance when needed. Does not affect  

performance. Always inspect shaft for damage.

Surfaces and Chamber

Cleaning

weekly

Maintains good lab practices for cleanliness. Always inspect for damage.

Most Common Error Codes and Clearing Procedures

ERROR CODE – 1 (tachometer error)

Clearing Procedure:

•  Wait for rotor stand still (120 seconds after rotor stops)

•  Turn the power off and wait 10 seconds

•  Turn the power on

•  Check to make sure rotor is installed properly 

 

  before spinning

ERROR CODE – 2 (mains interrupt)

Clearing Procedure:

•  Wait for rotor stand still

•  Open the lid

•  Press the start button

•  Close lid and start again

ERROR CODE – 3 (balance error)

Clearing Procedure:

•  Wait for rotor stand still

•  Check to make sure the centrifuge is sitting on a level   

  and sturdy surface

•  Make sure the rotor is installed properly and the bucket  

  weight is distributed evenly before spinning

ERROR CODE – d (lid lock error)

Clearing Procedure:

•  Turn the power off and wait 10 seconds

•  Turn the power on

•  Open and close lid

ERROR CODE – F (tachometer error)

Clearing Procedure:

•  Wait for rotor stand still

•  Turn the power off and wait 10 seconds

•  Turn the power on

•  Check to make sure rotor is installed properly 

 

  before spinning

Extract from the 

Operating Instructions

!