Sterilization
Sterilization capabilities, cleaning, disinfecting, handling, and storage of instrumentation
are the responsibility of qualified facility and/or user personnel. Devices must be
properly cleaned and disinfected, prior to sterilization.
After sterilization in the autoclave, let the accessory device cool down
slowly. NEVER use cold water to cool the motor. This will damage the
electronic components and seals.
Liquid on the cable connector of the accessory device can damage the
device. Before connecting the cable, ensure that the receptacles are clean
and dry
.
The control unit AR-200C and the foot control unit S-N1 are not approved
for sterilization.
7.1
Motor with Cable (AR-200M, AR-200M-ISO)
and Spray Clip (AR-200SP)
Sterilize the motor with cable, and the motor support and Spray Clip before
each use.
Please use dedicated cleaning adapters and tubing defined as accessories
in this User’s Guide for cleaning and sterilization of AR-200SP Spray Clip.
Approved Sterilization Procedures
Follow your country-specific directives, standards and guidelines.