Troubleshooting
DualWave Arthroscopy Pump
User’s Guide
LM1-0630-EN_E
Page 37 of 43
Message
Cause
** Pressure Fault **
1.
Ensure adequate fluid supply.
2.
Decrease outflow.
3.
Check tubing for damaged and if it is pinched, kinked or blocked.
4.
Check tubing for proper connections.
5.
Replace tubing.
6.
If failure remains, return to Arthrex for repair.
No Shaver Detect
1.
Ensure Universal shaver cable is connected.
2.
Turn On and Off DualWave while the shaver system is powered on.
3.
Adjust the DELAY and/or THRESHOLD settings.
4.
Isolate the power source.
5.
If failure remains, return to Arthrex for repair.
Troubleshooting Interference with Other Devices
11.1
Try one or more of the following to correct interference:
•
Reorient or relocate the receiving device.
•
Increase the distance between devices.
•
Connect the device to an outlet on a different circuit than the other
device(s).
•
Consult the manufacturer or field service technician for the receiving
device for assistance.