Setup
Note: Users may contact their Arthrex representative if they require clarification on
the Set Up instructions described below.
4.1
AC Power Safety Considerations
The AR-400UBC is powered by a medically rated universal AC input switching
power supply. This power supply allows users to connect the charger to any local
AC mains outlet.
Arthrex supplies separate region specific power cords (Hospital grade and EU) with
the AR-400UBC. Contact your Arthrex representative if you need a power cord that
must meet the electrical standards of another country.
Only use replacement power cords that comply with medical grade
standards, IEC 60320-1 Subclause 3.21, Detachable Power Supply
Cords or electrical standards for the designated country where the
AR-400UBC is being used.
Avoid positioning the unit so that it is difficult to disconnect the
coupler or plug from the supply main.
To prevent electrical shock, do not use extension cords or 2 to 3
pronged adaptors.
The charger has an AC mains switch on the rear panel. When the supply main switch
is OFF, no electrical power is used by the charger.
Two chargers may be connected via a the flex cable provided. No more than two
chargers may be connected in this way, otherwise the permitted electrical load values
may be exceeded and it could damage the units.
When the supply main switch is ON, the charger executes a series of self-diagnostic
tests. Upon successful completion of these tests, the charger will display the name
and model number. If the tests discover a problem, the problem is shown on the
display. Refer to Table 3 for a listing of AR-400UBC Display messages.
W A R N I N G !
To avoid the RISK of electric shock, this equipment must only be connected to a
SUPPLY MAIN with Protective Earth Terminal.