3.3.4
Cleaning the Camera Head
CAUTIONS:
•
If the camera head is dented or damaged, or if the cable silicone jacket is cut, DO NOT
autoclave or immerse in liquid (water, chemical disinfectants or sterilants, etc.). Notify
your Arthrex Sales Representative.
•
Do not place the camera head or accessories in an ultrasonic cleaner or ultrasonic
washer/sterilizer.
Preparation for Cleaning and Sterilization
Immediately after use, place the camera head assembly in a container and soak with neutral pH (PH 6.0 –
8.0) enzymatic cleaning solution (e.g. Enzol, Metrizyme or equivalent diluted to proper concentrations per
manufacturer’s instructions), in order to prevent blood, protein and other contaminants from drying onto the
camera head.
3.3.4.1
Automated Cleaning
•
Use only washers according to the International Standard ISO 15883.
•
Refer to the washer’s instruction manual.
1. Transfer the camera head into the washer for processing.
2. Make sure that the camera head has been securely fixed to the unit’s trays or baskets. Make sure
that the camera head does not touch other instruments.
3. Do not overload the washer.
4. Remove the camera heads immediately after the automatic procedure has stopped.
5. Set up and run washer for the wash cycle listed below.
Automatic Washer Cycle Definition
Phase
Recirculation Time
(Minutes)
Temperature
Detergent Type and
Concentration
Pre-Wash 1
02:00
COLD TAP WATER
NA
Enzyme Wash
03:00
HOT TAP WATER
Enzol® 1oz/gallon
Rinse 1
00:15
HOT TAP WATER 60
O
C
[140
O
F]
NA
Drying
06:00
90
o
C [194
O
F]
NA
6. Dry the equipment with a lint-free soft cloth. Wipe the card edge connector with 70% isopropyl
alcohol to remove any residual detergent.
a. Do not allow exposed glass windows to air dry. 70% isopropyl alcohol may be
applied to glass surfaces with a soft cotton applicator to prevent streaks and
spots. Dry the surfaces thoroughly with a cotton applicator after applying the
alcohol.
7. After cleaning, inspect the camera head assembly and camera head cable for cleanliness and
damage.
8.
CAUTION
: Inspect the camera head cable for breaks and cuts. Camera heads with damaged
cables should not be sterilized or disinfected. Return camera heads with damaged cables to
Arthrex for repair.
9. Before sterilization and/or disinfection, coil the camera head cable into loops at least six inches in
diameter. Do not kink or twist the cable.