3-11
You may exit the Output Screen without causing any changes to take effect by
touching the Close Button located at the upper right corner of the screen.
Saving a Case Log
The data for a current or any past case may be saved to an external USB
storage device.
1. Ensure that a supported USB storage device is connected to the USB Port.
Note:
Only devices or cables meeting IEC 60950 and IEC 60601-1 should be
connected to the USB Port and/or Ethernet Port (if present). Failure to do
so may result in operator shock. All cables used in conjunction with the
device should be no longer than 1 m (3 ft.) in length. Operators
connecting other devices to the USB Port must ensure compliance to the
system requirements of IEC 60601-1.
2. If the current case will not be saved, then select a past case to save as described
in “
Selecting Past Cases
” on page 3-9.
3. Touch the “
Output
” Button.
The Output Screen will appear on the
Angel System’s
Touch Screen Display
(see Figure 3-9).
The
Angel System
will then scan for a connected storage device, and the
text “
Searching for connected devices. Please wait…
” will appear on
the Output Screen.
If a USB storage device is detected, the “
Output log file to:
” Button will
become enabled.
4. Choose the appropriate destination by performing the following:
a.
Touch the “
Select destination for log file:
” field.
The field will become highlighted, and the Up and Down Arrow Buttons will
appear on the right side of the screen.
b.
Touch the Up and/or Down Arrow Buttons to scroll through available
destinations which the log may be saved to.
5. Once you have chosen your desired destination, touch the “
Output log file to:
”
Button.
A progress bar will appear to monitor the progress of the saving process. Once all data
lines have been saved, the text “
Output saved
” will be displayed beneath the “
Output
log file to:
” Button and it will return to the Output Screen.
If the Stop Button is pressed while the case is being saved, the save process will
stop, and an “
Output Failed!
” message will be displayed.
If the “
Output Tally to USB storage device
” Button or the “
Output log file to:
” Button
never becomes enabled, or if the “
Output failed!
” message appears when trying to
save, consult the “
Troubleshooting the Save Process
” on page 5-6.
Exit the Output Screen by touching the Close Button located in the upper right corner
of the screen. When you close this screen, it will return to the End of Case Screen or
the “
Tally
” tab.
Summary of Contents for Angel
Page 2: ...This page intentionally left blank ...
Page 8: ...Overview vi Angel cPRP System Operator s Manual This page intentionally left blank ...
Page 14: ...Overview xii Angel cPRP System Operator s Manual ...
Page 17: ...Overview Angel cPRP System Operator s Manual xv This page intentionally left blank ...
Page 18: ......
Page 26: ...Overview 1 8 Angel cPRP System Operator s Manual This page intentionally left blank ...
Page 66: ...Processing 3 28 Angel cPRP System Operator s Manual This page intentionally left blank ...
Page 98: ...Routine Care 6 12 Angel cPRP System Operator s Manual This page intentionally left blank ...