Output on/off and Tracking Operation Commands
Output[:STATe] {OFF|ON}
This command enables or disables the outputs of the power supply.
For example
:
OUTPUT ON Enable the outputs
OUTPUT OFF Disable the outputs
Output[:STATe]?
This command queries the output state of the power supply and returns the
corresponding values. “OFF” shows the output is disabled, and “ON” shows the
output is enabled.
Output Setting Commands
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] {<current>[MIN|MAX}
This command directly programs the immediate output current level of the output
selected with the INST command.
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]? [MIN|MAX]
This command checks and returns the immediate current value at the present output
terminals of the power supply or the maximum and minimum programmable current
levels of the selected output.
For example:
CURR? Query and return the immediate current value of the
output terminals
CURR? MAX Query and return the maximum programmable current
level of the output terminals
.
CURR? MIN Qurery and return the minimum programmable current
level of the output terminals
[SOURce:]CURRent[:LEVel]:LIMit[:AMPLitude] {<current>|MIN|MAX|DEF}
This command is to set the maximum current limit value of the output selected with
INST command.
[SOURce:]CURRent[:LEVel]:LIMit[:AMPLitude]? |MIN|MAX|DEF}
This query returns the current limit level of the presently selected output or the
minimum/ maximum/ default programmable current values of the presently selected
output.
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude] {<current>[MIN|MAX}
This command programs the pending triggered current level of the power supply,