AE-780 Operation Manual
Rev. 11/30/12 MKB
By connecting the 5
-
lb probes to the AE-780’s sockets, it is possible to measure
Resistance Point to Point (RTT), Resistance to Ground (RTG), and Volume Resistance.
Using these external probes will allow compliance with various standards including
ANSI/ESDA S4.1 for Work surface – Resistance Measurements, ANSI/ESDA S7.1
Resistive Characterization of Materials – Floor Materials.
When auditing is finished, unplug the cables and store probes in the protective case.
Resistance Point to Point (RTT)
RTT measurements can be used for the evaluation of floors, chairs, carts, work
surfaces and other ESD
-
controlled materials and products. Procedures vary regarding
sample preparation, probe preparation and spacing of the 5
-
lb probes. Select and read
the correct test procedure or standard for the desired measurement. Test procedures
can be purchased from
www.eastern-energy.com
A. Connect one end of each banana test lead into the sockets of the meter.
Connect the other end of the test coil cords into the 5
-
lb probes.
B. Place both probes on the material according to test procedures or standard being
used.
C. Press the “TEST” button and the value will be displayed on the LEDs. The
meter will apply the correct voltage (10v or 100v) according to the value of
what is measured.
D. When performing test, do not touch lead wires or probes. Avoid overlapping
of lead wires. This will ensure accurate readings.
Eastern Energy Co., Ltd.
Tel: 66-3844-6117
40/4 Vitoondumri Rd., Banbueng
Page 7 of 13