AM-BA SERIES HI-LO TREATMENT TABLES
Armedica Treatment Tables are intended to be used as physical therapy supports for patients
during clinician attended physical therapy related diagnosis, treatment, and monitoring. The table
is equipped with a vertical table height adjustment actuator and casters for mobile transport.
Patients are not to be left unattended.
INSTALLATION
1. Remove all packaging material from the tables.
2. There is no assembly required on ARMEDICA™ Treatment Tables.
3. Plug the power cord into a properly grounded 120 Volt AC outlet and follow the procedure
outlined in the Precautionary Instructions.
4. The table is equipped with four total lock swivel casters for maximum stability.
5. The motor is equipped with a thermal cutout that protects it from overheating.
PRECAUTIONARY INSTRUCTIONS
1. Read, understand and practice the precautionary instructions in this manual. Know the
limitations and the hazards associated with ARMEDICA™ Treatment Tables.
2. The motor is double insulated for protection from EMI. This equipment has been tested
and found to comply with the limits for medical devices to the IEC 60601-1-2: Third edition
2007-03. These limits are designed to provide reasonable protection against harmful
interference in a typical medical installation. This equipment generates, uses and can
radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to other devices in the vicinity. However,
equipment does cause harmful interference to other devices, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the interference
by one or more of the following measures:
- Reorient or relocate the receiving device.
- Increase the separation between the equipment.
- Connect the equipment into an outlet on a circuit different from that to which the
other device(s) are connected.
- Consult the manufacturer or field service technician for assistance.
3. Always make certain all four casters are locked prior to patient use.
4. Always use both hands when changing the angle of any section of the table.
5. Never change the angle of any section when the patient’s weight is on that section.
6. Never place your hands or feet nor the patient’s hands or feet near any of the working
mechanisms of the table when raising or lowering the table.
7. Never leave patient unattended.
OPERATING INSTRUCTIONS
To adjust the height of ARMEDICA™ Treatment Tables, a foot bar extending the length of
the table is used to activate the electric motor assembly. Pressing down on the bar will
raise
the table. Lifting up on the bar will
lower
the table.
Duty cycle on/off Int. 1 min./ 9 min.
TABLE SECTION ADJUSTMENTS
The sections of ARMEDICA™ Treatment Tables are raised by lifting them with both hands.
The locking mechanism holds them securely at any angle. To lower a section, hold it in one
hand and release the locking mechanism by turning the knob with the other.
CAUTION: When lowering any of the table sections, make certain that your hands are
placed where they cannot come in contact with the board supports.
CAUTION: When adjusting the sections with the patient on the table, be sure that the
patient’s weight is not on the section you are going to adjust
1