AM-BA SERIES HI-LO TREATMENT TABLES 

Armedica Treatment Tables are intended to be used as physical therapy supports for patients 

during clinician attended physical therapy related diagnosis, treatment, and monitoring. The table 

is equipped with a vertical table height adjustment actuator and casters for mobile transport. 

Patients are not to be left unattended. 

INSTALLATION 

 

 

1.  Remove all packaging material from the tables. 

 

 

2.  There is no assembly required on ARMEDICA™ Treatment Tables. 

 

 

3.  Plug the power cord into a properly grounded 120 Volt AC outlet and follow the procedure 

 

     outlined in the Precautionary Instructions. 

 

 

4.  The table is equipped with four total lock swivel casters for maximum stability. 

 

 

5.  The motor is equipped with a thermal cutout that protects it from overheating. 

 

PRECAUTIONARY INSTRUCTIONS 

 

 

1.  Read, understand and practice the precautionary instructions in this manual. Know the 

 

     limitations and the hazards associated with ARMEDICA™ Treatment Tables. 

 

 

2.  The motor is double insulated for protection from EMI. This equipment has been tested 

 

     and found to comply with the limits for medical devices to the IEC 60601-1-2: Third edition 

 

     2007-03. These limits are designed to provide reasonable protection against harmful 

 

     interference in a typical medical installation. This equipment generates, uses and can 

 

     radiate radio frequency energy and, if not installed and used in accordance with the 

 

     instructions, may cause harmful interference to other devices in the vicinity. However, 

 

     equipment does cause harmful interference to other devices, which can be determined by 

 

     turning the equipment off and on, the user is encouraged to try to correct the interference 

 

     by one or more of the following measures:  

 

 

 

- Reorient or relocate the receiving device. 

 

 

- Increase the separation between the equipment. 

 

 

- Connect the equipment into an outlet on a circuit different from that to which the 

   

 

  other device(s) are connected. 

 

 

- Consult the manufacturer or field service technician for assistance. 

 

 

3.  Always make certain all four casters are locked prior to patient use.                                                                                   

 

 

4.  Always use both hands when changing the angle of any section of the table. 

 

 

5.  Never change the angle of any section when the patient’s weight is on that section. 

 

 

6.  Never place your hands or feet nor the patient’s hands or feet near any of the working 

  

     mechanisms of the table when raising or lowering the table. 

 

 

7.  Never leave patient unattended

 

OPERATING INSTRUCTIONS 

 

 

To  adjust  the  height  of  ARMEDICA™  Treatment  Tables,  a  foot  bar  extending  the  length  of 
the table is used to activate the electric motor assembly. Pressing down on the bar will 

raise

 

the table. Lifting up on the bar will 

lower

 the table.    

 

 

Duty cycle  on/off   Int. 1 min./ 9 min. 

 

TABLE SECTION ADJUSTMENTS 

 

The  sections  of  ARMEDICA™  Treatment  Tables  are  raised  by  lifting  them  with  both  hands. 
The locking mechanism holds them securely at any angle. To lower a section, hold it in one 
hand and release the locking mechanism by turning the knob with the other. 

 

CAUTION:    When  lowering  any  of  the  table  sections,  make  certain  that  your  hands  are 
placed where they cannot come in contact with the board supports.  

 

CAUTION:    When  adjusting  the  sections  with  the  patient  on  the  table,  be  sure  that  the 
patient’s weight is not on the section you are going to adjust 

1

Summary of Contents for AM-BA Series

Page 1: ...red for the owners and operators of the ARMEDICA AM BA Series Hi Lo Treatment Tables It contains installation instructions precautionary instructions and maintenance procedures for the following model...

Page 2: ...to other devices which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate th...

Page 3: ...rotective earth connection The grounding pin located on the power supply cable is for functional use only and is present for EMI EMC purposes The product is Class II equipment and does not rely on thi...

Page 4: ...RL 4 4 4 4 4 4 24 02538 SCREW BTN HD SOC 1 4 20x3 4 2 2 2 2 2 2 25 02645 BOLT SHLDR 1 2x1 2 2 2 2 2 2 26 02646 BOLT SHLDR 1 2x1 1 4 2 2 2 2 2 2 27 02657 PLUG RD 1 ODx14 20 GA BLK 4 4 4 4 4 4 28 02660...

Page 5: ...AR FOOT SWITCH BC BASE PNTD 2 2 2 2 2 2 76 18958 BASE BA 4 CSTR 32 WIDE PNTD 1 1 1 1 77 18968 BASE BA 4 CSTR 37 1 4W PNTD 1 1 1 78 19021 PLATE FRMD SWITCH MNT 15 1 8 ASSEMB 1 1 1 79 19022 PLATE FRMD S...

Page 6: ...33 63 11 32 68 3 44 25 2 16 21 28 29 23 18 78 14 42 4 41 45 9 27 26 27 24 76 43 8 75 67 31 22 79 77 TABLE BASE AM BA 134 140 234 240 334 340 5...

Page 7: ...2 16 96 32 92 94 95 93 TOP FRAME AM 134 140 ITEM NUMBER 92 IS USED ON AM 134 ITEM NUMBER 93 IS USED ON AM 140 EXAMPLE REFER TO MODEL NUMBER COLUMN ON PARTS LIST FOR ITEMS WITH TWO REFERENCE NUMBERS 6...

Page 8: ...1 16 37 73 61 17 10 35 15 9 39 7 55 56 58 59 60 REFER TO MODEL NUMBER COLUMN ON PARTS LIST FOR ITEMS WITH TWO REFERENCE NUMBERS EXAMPLE ITEM NUMBER 55 IS USED ON AM BA334 ITEM NUMBER 58 IS USED ON AM...

Page 9: ...0 39 34 1 74 61 9 15 16 2 32 7 54 52 51 REFER TO MODEL NUMBER COLUMN ON PARTS LIST FOR ITEMS WITH TWO REFERENCE NUMBERS EXAMPLE ITEM NUMBER 54 IS USED ON AM BA234 ITEM NUMBER 52 IS USED ON AM BA240 TO...

Page 10: ...DISCARD RIVETS 2 Using both hands gently remove cover plate by pulling directly away from mounted switch plate 82 91 3 Use the foot control bar to determine which micro switch is not functioning then...

Page 11: ...BLUE BLUE BLUE PURPLE PURPLE GRAY GRAY GRAY GRAY GRAY PURPLE CORD TO MOTOR DOWN UP 86 90 89 82 85 83 80 84 87 81 88 91 98 99 97 OPEN CLOSED SWITCH BOX AM BA SERIES WIRE DIAGRAM 10...

Page 12: ...ble during the warranty period if the failure is not caused by a defect in material or workmanship or if the failure is caused by unreasonable use including the failure to provide reasonable and neces...

Reviews: