background image

1.4.1 Precautions for Instrument Installation

Before installation of the instrument, read the following notes and always take 

proper safety precautions.

• Determine a location for the instrument and install it in that location. Hold 

the bottom of the instrument with both hands when carrying it.

• Install the instrument on a surface with sufficient clearance around the 

instrument. Do not block the vents on the back of the instrument.

• Install the instrument where temperature and humidity can be maintained in 

the following ranges:

  Temperature: 10 - 35°C

  Humidity: 20 - 80%

  Installation in an environment outside these ranges may cause inaccurate 

measurement results.

• Install the instrument on a level, sturdy, vibration-free platform. Operation of 

the instrument in an unstable place may cause trouble with or malfunction of 

the instrument resulting in personal injury. Do not install the instrument 

where it may fall off or topple over.

• Do not use the instrument in a location where dust has accumulated. There 

is a risk of current leakage.

• Do not install the instrument near:

       - Places that store chemicals

       - Equipment that generates corrosive gas or electrical noise or

      - Equipment that may bring the instrument out of the operating  

        temperature and humidity ranges.

These factors may cause trouble with or malfunction of the instrument, 

resulting in personal injury, or may otherwise cause inaccurate measurement 

results.

• Install the instrument in a place where condensation, intense light such as 

direct sunlight or wind can be avoided. These factors may cause inaccurate 

measurement results, as well as deformation of or damage to the instrument.

• Apply the correct voltage (100 - 240 V AC ± 10%) and frequency (50/60 Hz) 

to the instrument. Improper voltage and frequency may result in fire or 

damage to the instrument and consequently lead to personal injury.

• Use the power cord and AC adapter that come with the instrument. 

• Connect the AC adapter directly to a single outlet. The power supply for the 

instrument is 33 VA. If you use a power strip, make sure it is grounded. Also, 

check if the power capacity for the power strip is within the specified range.

• It is possible to use a commercially available mobile battery in place of the 

provided AC adapter. Use a battery with a rating of 5 V and 2.0 - 2.4 A.

• Do not disassemble the instrument. Do not modify the instrument. 

Disassembly and modification of the instrument may result in exposure to 

pathogenic microorganisms or cause fire or damage to the instrument and 

consequently lead to personal injury.

1.4.2 Precautions for Instrument Relocation

Before relocating the instrument, read the following notes and always take 

proper safety precautions.

• Hold down the      button to turn off the power, and then disconnect the 

power cable from the instrument.

• Remove the reagent strip from inside the instrument. There is a risk of 

pathogenic microbes scattering inside the instrument if the instrument is 

moved with a reagent strip still inside.

• If the instrument will be moved a long distance, secure the strip tray and 

pipette guide with tape to ensure that neither opens or comes off.

• There is a risk of instrument failure if moved without securing the strip tray 

and pipette guide. Place the instrument in the dedicated box together with 

the accessories to move.

• Do not subject the instrument to shock or vibration when moving. There is a 

risk of instrument failure.

• Before moving the instrument, read “1.4.1 Precautions for Instrument 

Installation”.

1.4  Installation

1.2  Unpacking

1.3  Part Names and Functions

1.3.1  Front View

1.3.2  Rear

1.1.3  Measurement Principle

■ Reagent strip measurement

The instrument is equipped with multiple light source LEDs of varying 

wavelength. The light is dispersed on the light path and transmitted to the 

photometric unit. The dispensed sample fills the photometric cell and reacts 

with the reagents in the cell, changing the absorbance and turbidity. The light 

source LED shines on the cell, and the light that transmits through is read by 

a photodiode (receiver). The amount of transmitted light that is read is 

converted to a concentration.

Unpack the box and make sure you have all items listed in this section. If 

anything is missing or damaged, contact your distributor.

• Instrument

• Power cord

• AC adapter

• Operating manual

• Pipette

NOTE:

The following items are not included with the instrument: 

mobile battery, disposable gloves, tips, reagent strips, control, cotton 

swabs, distilled water, soft cloth, alcohol.

The alcohol shown in this manual means 70% isopropyl alcohol.

c

 Pipette guide

    attachment rails

a

 Display

b

 Operation 

     buttons

d

 Strip tray

e

 IC-CHIP 

     insertion port

f

 USB port

g

 Power input 

     terminal

Summary of Contents for Lab 004

Page 1: ...in close proximity to sources of strong electromagnetic radiation as these may interfere with proper operation Premise Symbols Read this operating manual thoroughly before using the instrument This operating manual gives an overview of the instrument and the proper procedures for operation and maintenance Follow the instructions in this operating manual in order not to defeat the purpose of the pr...

Page 2: ...d in package insert of reagent Measurement principle Listed in package insert of reagent Dual wavelength transmitted light measurement Number of reagent strips measurable at once 1 Measurement time Listed in package insert of reagent Approximately 4 μL Sample consumption Listed in package insert of reagent 1 sample Approximately 10 minutes in environments 25 C and below Memory capacity 50 measurem...

Page 3: ...adapter Use a battery with a rating of 5 V and 2 0 2 4 A Do not disassemble the instrument Do not modify the instrument Disassembly and modification of the instrument may result in exposure to pathogenic microorganisms or cause fire or damage to the instrument and consequently lead to personal injury 1 4 2 Precautions for Instrument Relocation Before relocating the instrument read the following no...

Page 4: ... infectious diseases Improper handling of blood may cause infection to the user or other individuals by pathogenic microorganisms Dispose of used samples in accordance with local regulations for biohazardous waste When using a sample that has been stored in a refrigerator or freezer allow the sample to acclimate to the ambient temperature 10 35 C prior to measurement Do not store a sample in direc...

Page 5: ... from a New Box Wear disposable gloves to prevent exposure to pathogenic microorganisms If sample adheres to the instrument wipe it off with a cotton swab or gauze dampened with distilled water Make sure that no dirt is left and then ensure that the surface is disinfected with alcohol Leaving the instrument with the sample adhered to it may cause infection to the user or other individuals by patho...

Page 6: ...d time are set a On the standby screen press and hold for 3 seconds to display the MENU screen 1 0 0 will appear in the sub screen with the 1 digit blinking b Press the button The 1 will stop blinking and the middle 0 digit will change to 1 and start blinking c Press the button once The middle digit will change to 2 d Press the button The MENU No will be finalized as 1 2 and the screen will transi...

Page 7: ...red Use a new reagent strip and repeat measurement E 124 The tray was withdrawn during measurement Press the button to clear the error and repeat measurement E 126 The measurement count limit of the inserted IC CHIP has been exceeded Insert the IC CHIP inside of the box of the reagent strip to be measured into the port on the back of the instrument and repeat measure ment E 128 Unable to read the ...

Page 8: ...omatic dry biochemical analyzer RATING MADE IN PHILIPPINES Environment During Transport Semi Auto matic dry biochemical analyzer RATING MADE IN PHILIPPINES Storage Environment Storage Environment Environment During Transport This product conforms to Regulation EU 2017 746 In vitro diagnostic medical device Manufacturer Date of manufacture Authorised Representative in the European Community Importe...

Reviews: