70
Warnings, Precautions & Limitations
• Never make significant changes to your diabetes control program or ignore
physical symptoms without consulting with your healthcare professional.
• Do NOT use the “GLUCOCARD™ X-METER” to measure the blood glucose
level of patients who have received treatment containing maltose,
icodextrin, who have undergone a galactose tolerance test, or xylose
absorption test. This system may overestimate their blood glucose level and
may thus mask hypoglycemia.
• Icodextrin and its metabolites (maltose, maltotriose, and maltotetraose) will
cause falsely higher results.
• Do not use during a xylose absorption test for diagnostic evaluation of
malabsorption as this will cause falsely higher results.
• Infusion of galactose will cause falsely higher results.
• EDTA, heparin, and mono-iodoacetic acid samples of blood may be used.
Do not use other preservatives.
• Do not grasp the meter near the test strip port. A built-in thermosensor used
to minimize measurement errors is installed just inside the port. If you hold this
area, it will be warmed by your fingers, which will adversely affect test
results.
Summary of Contents for Glucocard X-Meter
Page 1: ...USER INSTRUCTION MANUAL ...