67
Your blood glucose level is less than 20 mg/dL or there
was not enough blood on test strip. Repeat test with
new test strip. Make sure you use enough blood to
fill the strip. If this message shows again, contact
your healthcare professional immediately!
Your blood glucose level is more than 600 mg/
dL. Repeat test with new test strip. If this message
shows again, contact your healthcare professional
immediately.
Battery power is too low for testing. Replace batteries
immediately. Results may not be accurate when
testing with low battery symbol on.
Thermometer symbol. The meter temperature is
outside acceptable range of 50˚F-104˚F (10˚C-40˚C).
Allow the meter and test strips to warm up or cool
down slowly (20 minutes) until the temperature
warning symbol goes away.
Summary of Contents for GLUCOCARD 01-mini
Page 1: ...User Instruction Manual ...