Chapter 1 Before Using the AE-4022
AE-4022
1-12
1.3.2
Installation of the instrument
Certain components are held securely using fixing tape to protect the instrument from damage during transportation.
The tapes must be removed and these security measures released to install the system and prepare the instrument for
use.
Before installation, read “1.3.1. Precautions for installation” on page 1-11.
Items required: AC adapter, power cord, carrying arm, and connection cable (sold separately) when communicating
with an external device.
1
Release the carrying arm and maintenance cover.
a
Remove the fixing tapes from the carrying arm and maintenance
cover.
2
Open the maintenance cover
a
Press and hold the two cover open buttons on either side of the
instrument at the same time (
(1)
in the right figure) and open the
maintenance cover (
(2)
in the right figure).
Use an RS-232C cable to connect an external device to the instrument. Connection using
other than an RS-232C cable can cause electric shock and fire. For more details, contact your
local distributor.
Fixing tapes
Cover open
button
(1)
(1)
(2)