1.3 Installing the instrument
AE-4022
1-11
1.3
Installing the instrument
1.3.1
Precautions for installation
Before installing the instrument, read the following items and always take proper safety precautions.
Install the instrument at least 20 cm away from walls. Inadequate clearance between the
instrument and walls may cause overheating of the instrument or undesirable load on cable
connections, thus resulting in fire or incorrect measurement results.
Apply the correct voltage to the instrument. The wrong voltage may result in fire or damage
to the instrument and consequently lead to personal injury.
Do not
disassemble or modify the instrument without our approval. This may result in fire,
damage to the instrument and consequently personal injury.
Install the instrument on a level, vibration-free sturdy platform. Operation of the instrument in
an unstable place may cause trouble or malfunction of the instrument resulting in personal
injury.
Do not
install the instrument where it may fall off or topple over.
Do not
install the instrument near places that store chemicals or near equipment that
generates corrosive gas or electrical noise. Chemicals, corrosive gases and electrical noise
may cause fire or damage to the instrument and consequently lead to personal injury, or
may otherwise cause incorrect measurement results.
Install the instrument in a place not exposed to direct sunlight, condensation or wind. These
factors may cause incorrect measurement results, as well as deformation of or damage to
the instrument.
Use an RS-232C cable to connect an external device to the instrument. Connection using
other than an RS-232C cable may cause electric shock and fire.
Use an external communication cable shorter than 30 m. External communication cables
are for indoor use only.
Install the instrument where temperature and humidity can be maintained in the following
ranges.
Temperature: 10°C - 30°C
Humidity: 20% - 80% R.H.
Installation in measurement environments outside these ranges may result in incorrect
measurement results.