Arkray AUTION ELEVEN AE-4022 Quick Reference Manual Download Page 1

SEMI-AUTOMATED URINE ANALYZER

AUTION ELEVEN 

AE-4022

Quick Reference Guide

Summary of Contents for AUTION ELEVEN AE-4022

Page 1: ...SEMI AUTOMATED URINE ANALYZER AUTION ELEVEN AE 4022 Quick Reference Guide ...

Page 2: ...e Waste Box 11 Replacing Thermal Paper 12 Warnings 13 Errors 14 15 This Quick Reference Guide is a brief summary of the operation and routine maintenance of the AUTION ELEVEN Semi Automated Urine Analyzer Read the Operating Manual before using the instrument Wear protective gloves to prevent exposure to pathogenic microbes SEMI AUTOMATED URINE ANALYZER AUTION ELEVEN AE 4022 ...

Page 3: ...System Overview System Overview ...

Page 4: ...strip tray Carrying arm Cover open button on both sides Waste box Maintenance cover Built in printer Rails white Test strip detection window Preparation of measurement Setup Meas Check that there is no used test strips in the waste box S Select t Dip the Set the Place a strip de of the te The car analyze Set the System Overview ...

Page 5: ...Sample Preparation Sample Preparation ...

Page 6: ...no used test strips in the waste box Turn on the power switch at the back of the instrument Prepare a sufficient volume of sample in urine collection cup so that the entire test strip pad area can be dipped in a single motion Stir well Waste box Test strip Test strip pad area 2 3 Preparation for measurement bPress The recorded patient ID number is stored and the display returns to the Standby scree...

Page 7: ...Sample Measurement Sample Measurement ...

Page 8: ...umber will display alternately on screen AE 4022 2 12 1 Call up the screen aWith the Standby screen displayed press and hold for 3 seconds The Operator ID number input screen appears 2 Enter your operator ID number aEnter your operator ID number correctly using the numeric and keys bPress If your entry is valid the display returns to the Standby screen You are then permitted to perform measurement...

Page 9: ...lection cup Set the measurement number Place a test strip over the detection area of the test strip detection window crossing the two white rails of the test strip tray The carrying arm sends the test strip into the analyzer and the measurement starts automatically Set the patient ID Do NOT wet the black mark Test strip Test strip detection window Detection up to 13 digits up to 4 digits IMPORTANT...

Page 10: ... mark After dipping the test strip in the sample remove excess urine by letting the strip touch the edge of the collection cup Place a test strip over the detection area of the test strip detection window crossing the two white rails of the test strip tray The carrying arm sends the test strip into the analyzer and the measurement starts automatically When measuring continuously repeat steps 1 to ...

Page 11: ...Control Measurement Control Measurement ...

Page 12: ...trip in the sample for 2 seconds and then withdraw it IMPORTANT Dip the entire test strip pad area in the sample at once Do not allow the black mark shown in the figure to become wet If the test strip is dipped icon is displayed 2 Enter measurement number to identify control level 3 Dip entire test strip pad area in sample for two 2 seconds If buzzer setting is ON follow the timing chart below AE ...

Page 13: ...Check Measurement Check Measurement ...

Page 14: ... parameter See Operational mode on page 2 3 The standard setting is Auto start mode To change the operational mode when turning ON see 3 7 8 No 005 Operational mode when turning ON on page 3 8 When using the Cycle start mode press to start the check measurement of the first strip In the Auto start mode the system automatically starts the check measurement of the first strip when placement of the c...

Page 15: ...Maintenance Maintenance ...

Page 16: ... any dirt Carrying arm Suction port Waste box Test strip feed tray Feed lever Test strip tray Wipe the parts with cloth and fully dry them before attaching them back on This Quick Reference Guide is a brief summary of the measurement operation and daily maintenance Before using the instrument make sure to read the Operat Clean the area circled in red below Suction port Detach the suction port Test...

Page 17: ... feed lever by sliding the white lever in the center towards you Wash the detached carrying arm waste box test strip tray suction port test strip feed tray and feed lever with water to remove any dirt Carrying arm Suction port Waste box Test strip feed tray Feed lever Test strip tray Wipe the parts with cloth and fully dry them before attaching them back on This Quick Reference Guide is a brief su...

Page 18: ...ve any dirt Carrying arm Suction port Waste box Test strip feed tray Feed lever Test strip tray Wipe the parts with cloth and fully dry them before attaching them back on This Quick Reference Guide is a brief summary of the measurement operation and daily maintenance Before using the instrument make sure to read the Operat ing Manual Clean the area circled in red below r Wash the detached carrying...

Page 19: ...with alcohol then wash with water and wipe dry 3 Attach waste box 2 4 10 Items required Alcohol cloth and protective gloves 1 Discard used test strips aCheck that the Standby screen is displayed bPull out the waste box and discard the used test strips 2 Sterilize and clean the waste box aSterilize the waste box using alcohol and then wash with water to remove all traces of dirt bWipe all moisture ...

Page 20: ...ly run out remove the paper core from the paper holder and go to step 3 2 Remove the remaining thermal recording paper aPress The thermal recording paper remaining in the printer is fed out Remove the paper by holding it with your fingertips Printer cover Take out Cut using scissors button after replacing the thermal recording paper If you do not fully feed the thermal recording paper through the ...

Page 21: ...Warnings Warnings ...

Page 22: ...ult is not obtained When a problem occurs during measurement the system continues measurement and prints warning messages with the measurement results The table below explains the meaning of and measures corrective action to be taken for each warning message Chapter 5 Troubleshooting 5 1 Warning messages A warning message appears when a normal measurement result was not obtained When a problem occ...

Page 23: ...Errors Errors ...

Page 24: ... of thermal recording paper in the paper holder orienting it correctly so that the paper feeds from the bottom Guide the trimmed end of the thermal recording paper into and through the slot that receives the recording paper The thermal recording paper is automatically wound forward and fed into the unit NOTE When the thermal recording paper is completely removed from the printer the screen message...

Page 25: ...ty and the error recurs contact your local distributor E009 No test strip on the feeder The feed lever is not attached The incoming strip sensor is faulty The sensor failed to detect a test strip Turn OFF the instrument and open the maintenance cover Mount the feed lever If you find any obstacles or scattered test strips inside remove them If you find no abnormality and the error recurs contact your...