MAGX2 User Guide
25
V2.14 1-2-2017
3.5.14. User Settings > Frequency output
Back with no change
Item selection
Enter this part of the menu
User Settings > Frequency output > Settings Signal
This function serves to select which signal the output should be
giving.
Back with no change
Item selection
Selection identification
Confirmation of setup and saving to memory
OFF
Output:
OFF
Fixed
Output:
100Hz
fixed output
Flow+
Output:
100Hz
, for any positive flow
Flow
–
Output:
100Hz
, for any negative flow
Error
Output:
100Hz
, for any error identified by the device
Air Detect
Output:
100Hz
, during air detection (empty pipe)
Direct
Driving
Frequency output changing according to actual flow
User Settings > Frequency output > Direct Driving
This function serves to set flow values in relation to frequency output.
Possible range:
0
– 36000 m3/h
;
0
– 1000 Hz
Back with no change
Selection of digit position
Value setting
Confirmation of setup and saving to memory
Flow
min.
– max.
Setup of the active flow-range for the Frequency
output module
F min.
– max.
Configuration of the Frequency output range,
corresponding to the actual flow-rate within the given
range.
Minimal frequency 2 Hz
Maximal frequency 1000 Hz