31
GROUND RESISTANCE
Range
0.1 Ω to 10.0 Ω
Tolerance
± 1% + 0.5 ohm
INSULATION RESISTANCE TEST
Range
0 MΩ to 200 MΩ (applied voltage - 500VDC)
Tolerance
± 1%
LEAKAGE CURRENT
Range - Tolerance
•
0μA - 500μA (DC to 1kHz - ± 1%)
•
500μA - 1mA (1kHz to 100kHz - ± 2,5%)
• 1mA - 2mA (100kHz to 1MHz - ± 5%)
• 2mA - 12mA (100kHz to 1MHz - ± 5%)
FUNCTIONAL CURRENT
Range
0.1 A to 10.0 A*
*limited to protection fuse
Tolerance
± 5%