INTRODUCTION
SAFETY INFORMATION
LUFT
Mechanical Ventilation Analyzer allows the verification of
Anesthesia Devices, Lung Ventilators, Oxygen Concentrators,
Flowmeters, Pressure Manometers of Medical Gas Networks, among
other applications. Luft measures bidirectional, unidirectional, high-
and low-pressure airflow, barometric pressure, oxygen concentration,
and ambient temperature and humidity.
It is integrated with the management system, enabling real-time integration
and reading of the values tested, registering them in the device’s repair
and maintenance records. Luft operates on a rechargeable lithium-ion
battery, capable of lasting three to five hours of uninterrupted testing.
For better conservation of your device and to prolong its useful life,
please consider the following:
1.
Do not expose the device to high humidity.
2.
Do not expose the device to temperatures above 50º C.
3.
Respect the device’s battery life.
4.
Insert the connectors only into the spots intended for operation.
5.
Do not connect this device to the patient.
6.
Replace the power cord if any damage is identified.
7.
Replace accessories if any damage is identified.