792-681-07
9
WARNING
The procedures below must be carried out by suitably trained and qualified personnel. Failure
to do so may result in injury or an unsafe product.
Actions to be done by qualified personnel
Every six
months
Carry out a test of electrical couch positioning functions.
9
Check operation of the castors.
9
Examine the power supply cord and mains plug; if damaged, replace the complete
assembly; do not use a rewireable plug.
9
Examine all accessible flexible cables for damage and deterioration.
9
Check all accessible nuts, bolts and other fasteners are present and correctly
tightened.
9
Check hydraulic couches for smooth operation and signs of leakage.
9
Check any accessories fitted to the couch, paying particular attention to fasteners and
moving parts.
9
WARNING
All electrical checks should be performed by a suitably qualified and licensed electrician.