![Arjohuntleigh Sara Lite Instructions For Use Manual Download Page 23](http://html.mh-extra.com/html/arjohuntleigh/sara-lite/sara-lite_instructions-for-use-manual_2972576023.webp)
23
Technical Specifications
0
0
1
.20
05
8.E
N
re
v.
10
PRODUCT INFORMATION
Weight, complete
59 kg (130 lb)
Lifting capacity (SWL)
175 kg (385 lb)
Battery pack weight
5 kg (11 lb)
Minimum door requirement
660 mm (26 in)
Operating Force of Control
< 21 N
ELECTRICAL
Degree of protection of hand control
IPX7
Internally powered
24 VDC
Duty cycle
10%, 6 min / hour, 1 min continuous
Sound power level
< 65 dBA
Degree of protection (hoist)
Refer to product label
Fuse - Overload
5 A (thermal cutout)
Fuse - PCB
10 A (time delay)
Medical equipment
Refer to product label
The SARA LITE conforms to ISO 10535: 2006, AAMI ES60601-1,
CSA C22#60601-1, IEC 60601-1, CAN/CSA C22.2 No 601.1-M90, UL 60601-1.
WARNING: Wireless communications equipment such as wireless home network devices, mobile
phones, cordless telephones and their base stations, walkie-talkies, etc., can affect the SARA
LITE and should be kept at least 2.34 m away from it. Cables from potentially strong sources of
electromagnetic fields should not be placed near the unit. See the “Electromagnetic
Compatibility”section for more details.
BATTERY AND CHARGER UNIT
Battery type
Rechargeable (sealed lead acid)
Battery capacity
24 V, 4 Ah
Battery charger (part number NDA8200) - Input
100-240 VAC, 50-60 Hz, 50 VA
Battery charger rated output voltage
24 VDC, 24 VA, 1 A
Battery charger - protection class - intrusion of liquid
IP40
Battery charger - protection class - shock prevention
Class 2, double insulated
OPERATION AND STORAGE CONDITIONS
Ambient temperature range
Operation: 5° to 40°C (+41° to +104°F)
Storage : - 25° to 70°C (-13° to 158°F)
Relative humidity range
Operation: 15 % to 93 %, non condensing
Storage : 0 % to 93 %, non condensing
Atmospheric pressure range
Operation: 795 hPa to 1060 hPa (2000 m max)
Storage : 500 hPa to 1060 hPa
WARNING: This equipment is not suitable in the presence of flammable anesthetic mixtures with
air or oxygen, or with nitrous oxide. Using the SARA LITE in this environment might lead to an
explosion. The lift might create some spark internally and ignite the gaz.
RECYCLING
Battery
Sealed lead-acid, rechargeable, recyclable
Package
Cardboard recyclable
The lift
Separated and recycled, according to the European Directive
2002/96/EG (WEEE).
Technical Specifications