Charging the Battery

General

• The charger converts mains voltage to low 

DC voltage.

• The charger must only be used for

 

Arjo-

Huntleigh battery 

NDA

0100.

• The charger is ready for use as soon as it is 

plugged in.

• Disconnect the mains by unplugging the 

charger.

• A orange indicator LED indicates that the 

charger is ready for charging, there is no 
On/Off button on the charger.

Charging intervals

• To prolong life of batteries, charge at regu-

lar intervals.

• If the ArjoHuntleigh product is used fre-

quently, its recommended to use two batter-
ies, one in operation and one being charged.

WARNING

Do NOT charge the battery in a sealed 
compartment due to fire hazard.

CAUTION

To avoid breaking the charger when 
unplugging the battery connector: 
Grip the plug and don’t pull the cord.

NOTE

The battery can be left in the charger unit 
even if it is fully charged without being 
overcharged.

Charging procedure

1 Plug in the charger in a power outlet
2 Attach the battery to the charger.

3 The mains indicator LED will stay orange 

during the charge.

4 Once the battery is charged the orange 

charge indicator LED will turn from orange 
to green.

Charge
Indicator 
LED

Summary of Contents for NDA 1200

Page 1: ...with people in mind 04 ND 02_2GB September 2014 NDA 1200 2200 4200 6200 Battery Charger INSTRUCTIONS FOR USE...

Page 2: ...oup of companies Arjo Hospital Equipment AB 2012 As our policy is one of continuous improvement we reserve the right to modify designs with out prior notice The content of this publication may not be...

Page 3: ...ations 7 Installation of Charger 8 Charging the Battery 9 Care and Preventive Maintenance 10 Troubleshooting 12 Technical Specifications 13 Labels on the Battery Charger 14 Electromagnetic Compatibili...

Page 4: ...equipment and other appliances supplied by ArjoHuntleigh to avoid injuries caused by the use of inadequate parts Unauthorized modifications on any Arjo Huntleigh equipment may affect its safety ArjoH...

Page 5: ...charger is intended for charging Arjo Huntleigh battery NDA 0100 only in accord ance with the instructions outlined in the Instructions for Use IFU All other use must be avoided Installation Service...

Page 6: ...th the battery Should the battery casing crack and cause contents to come in con tact with skin or clothing rinse immediately with plenty of water If contents come in contact with the eyes rinse immed...

Page 7: ...at all times 8 Make sure to have a rescue plan ready in case of a resident emergency 9 If you have any questions please contact your local ArjoHuntleigh representative for support and service The cont...

Page 8: ...d in bathrooms or any other wet areas IP2X Mount the charger to a suitable wall with four screws and plug it in to a standard power AC outlet For screw specifica tions see Technical Specifications in...

Page 9: ...oduct is used fre quently its recommended to use two batter ies one in operation and one being charged WARNING Do NOT charge the battery in a sealed compartment due to fire hazard CAUTION To avoid bre...

Page 10: ...and standards may be more stringent than the recommended maintenance schedule CAREGIVER OBLIGATIONS Action Check Every WEEK Visually check all exposed parts X Visually check battery and charger X WAR...

Page 11: ...11 Visually check battery and charger Visually check the condition of the Battery Charger its cable and the battery Withdraw and replace immediately if damaged...

Page 12: ...ROBLEM ACTION The LED does not light up 1 There is no electricity coming from the AC outlet 2 The charger is broken The charger and or power supply cord is bro ken Order a new charger NEVER open the c...

Page 13: ...for mounting the charger Nrd 90N Resistance tensile Vrd 250N Resistance shear Mains Disconnection device Mains disconnection device must be visible and accessible at all times Disconnect mains by unp...

Page 14: ...voltage and charger current NDA0100 ArjoHuntleigh battery intended for this charger Insulation Class II Read the IFU before use IP2X Use in dry areas CE marking in accordance with The Medical Devices...

Page 15: ...15 Identification label on the Battery Charger Example of data label US Example of data label on battery holder Example of label on package...

Page 16: ...he location Guidance and manufacturer s declaration electromagnetic emissions The ArjoHuntleigh Battery Charger is intended for use in the electromagnetic environment specified below The customer or t...

Page 17: ...EC 61000 4 4 2 kV for power supply lines 1 kV for input output lines 2 kV for power supply lines input output lines not applicable Mains power quality should be that of a typical commercial or hospita...

Page 18: ...61000 4 6 Radiated RF IEC 61000 4 3 3 Vrms 150 kHz to 80 MHz 3 V m 80 MHz to 2 5 GHz 3 Vrms 3 V m Portable and mobile RF communications equipment should be used no closer to any part of ArjoHuntleigh...

Page 19: ...19 Parts and Accessories Battery Charger NDAX000 XX Extra battery NDA 0100...

Page 20: ...20 Intentionally left blank...

Page 21: ...21 Intentionally left blank...

Page 22: ...22 Intentionally left blank...

Page 23: ...582 745 866 E mail international arjohuntleigh com ITALIA ArjoHuntleigh S p A Via di Tor Vergata 432 IT 00133 ROMA Tel 39 0 6 87426211 Fax 39 0 6 87426222 E mail Italy promo arjohuntleigh com NEDERLAN...

Page 24: ...oHuntleigh GETINGE and MAQUET ArjoHuntleigh focuses on patient mobility and wound management solutions GETINGE provides solutions for infection control within healthcare and contamination prevention w...

Reviews: