21
5. Decontamination
The following processes are recommended, but should be adapted to
comply with the local or national guidelines (Decontamination of
Medical Devices) which may apply within the Healthcare Facility or the
country of use. If you are uncertain, you should seek advice from your
local Infection Control Specialist.
The
Lifeguard
system should be routinely decontaminated between
patients and at regular intervals while in use; as is good practice for all
reusable medical devices.
WARNING
Protective clothing should always be worn when carrying out
decontamination procedures.
Caution
Do not use Phenol-based solutions or abrasive compounds or
pads during the decontamination process as these will damage
the surface coating. Do not boil or autoclave the cover.
To Clean
Clean all exposed surfaces and remove any organic debris by
wiping with a cloth moistened with a simple (neutral)
detergent and water. Dry thoroughly.
Chemical
Disinfection
To protect the integrity of the cover we recommend a
chlorine-releasing agent, such as sodium hypochlorite, at a
strength of 1,000ppm available chlorine (this may vary from
250ppm to 10,000ppm depending on local policy and
contamination status).
Wipe all cleaned surfaces with the solution, rinse and dry
thoroughly.
Alcohol based disinfectants (maximum strength 70%) may
be used as an alternative.
Ensure the product is dry before storage.
If an alternative disinfectant is selected from the wide variety
Summary of Contents for Lifeguard LG20
Page 1: ...Lifeguard LG20 LG50 Instructions for Use with people in mind 746 158 08CN REG 04 2016 ...
Page 2: ...Lifeguard LG20 LG50 Instructions for Use with people in mind 746 158 UK_08 10 2015 ...
Page 30: ...Lifeguard LG20 LG50 使用说明书 with people in mind 746 158 CN_08 02 2016 ...
Page 32: ...1 目录 1 前言 3 2 临床应用 4 3 操作 5 4 配件 19 5 清洗消毒 21 6 保养和防护性维护 23 7 保修和服务 25 8 尺寸与数据 26 ...
Page 40: ...9 LG20 高度调节 升高平台 通过抽吸动作操作脚踏板 A 降低平台 升高踏板 A 或踩踏踏板 B 并将其一 直踩住 当平台处于所需高度或者达到最低位置 时 释放踏板 ...
Page 58: ...Intentionally left blank ...