20
4
Momentarily press DOOR UP.
The door should move smoothly to the position of
tub inner bottom (the seal should not be seen).
If the door during this procedure moves backwards under
the tub without any visible reason, check with an ARJO
service technician.
B2. The optional Lift-Accessible door function
- test of jam protection
1
Push and hold the DOOR DOWN button and the
door shall run smoothly down.
2
Let the door stop in the position of the inner bottom
(the seal should not be seen).
3
Momentarily press DOOR DOWN.
The door should move smoothly in under the tub.
The motor shall stop without noise when the door
reaches about 450 mm (18") in under the tub.
If this does not happen, the tub must not be used
before inspection by an ARJO service technician.
4
Momentarily press DOOR UP.
The door should move smoothly to the position of
tub inner bottom. Before this position, put a pencil
between the door and the seal profile to prevent con-
tinued movement. The door
shall move backwards
when the pencil is hit. If this does not happen, con-
sult an ARJO service technician.
•
Check the door lock
1
With the right cover removed, run the door to top
position. Observe from the side of the tub, the posi-
tion of the door lock in relation to the door profile.
2
The position should be according to the picture
below. If adjustment is needed, consult an ARJO
service technician.
•
tight, but no contact