33
Miscellaneous
EMC
first step select
™
mrs uses rf energy only for its internal function. therefore, its rf emissions are very low and are not likely to
cause interference in nearby electronic equipment. the first step select
™
mrs was tested to the following standards.
•
EN 60601-1-2: 2007
•
Conducted Emissions: Nov. 21, 2008 (Group 1, Class B)
•
Radiated Emissions: Dec. 17, 2008 (Group 1, Class B)
•
EN 61000-4-2: Dec. 3, 2008 (6/8 kV)
•
EN 61000-4-3: Dec. 1, 2008 (3 V/m)
•
EN 61000-4-4: Dec. 18, 2008 (2 kV)
•
EN 61000-4-5: Dec. 1, 2008 (1 kV)
•
EN 61000-4-6: Dec. 24, 2008 (3 Vrms)
•
EN 61000-4-8: Dec. 25, 2008 (3 A/m)
•
EN 61000-4-11: Dec. 21, 2008 (>95%-0.5p, 30%-25p,>95%-250p)
•
EN 61000-3-2: Dec. 21, 2008 (Class A)
•
EN 61000-3-3: Dec. 24, 2008
Classification
Classification in accordance with UL 60601-1, CSA C22.2 No 601.1 and IEC. 60601-1
•
Class II (UL 60601-1)
•
type B
•
no sterilization
•
not for Use with flammable anesthetic mixture with air, oxygen or nitrous oxide.
•
Continuous operation
•
ordinary equipment
Environmental Storage / Transit Conditions
Temperature Range: -20˚ to 60˚ C (-4˚ to 140˚ F)
Relative Humidity Range: 10% to 95%
Atmospheric Pressure Range: 500 hPa to 1060 hPa
Environmental Operating Conditions
Temperature Range: 10˚ to 40˚ C (50˚ to 104˚ F)
Relative Humidity Range: 10% to 95%
Atmospheric Pressure Range: 500 hPa to 1060 hPa