1
Power on indicator
2
Battery indicator
3
Raise / lower mattress platform height
4
Raise / lower backrest
5
Raise / lower thigh section
6
Auto-Chair (described below)
7
Head down tilt (Trendelenburg)
8
CPR - hold down the button to return the bed to a
ÀDWSRVLWLRQ
9
Foot down tilt (Reverse Trendelenburg)
10
Function lock / unlock
11
Bio-Contour (described below)
Patient Controls
Caregiver Controls
Patient Handset
Attendant Control Panel (ACP)
Battery status indicator
2
Battery fully charged
(indicator off)
Battery charging
Battery discharged
Bed controls
The Auto-Chair up button simultaneously raises
the backrest and thigh sections, pausing when the
backrest reaches 45º. Continue to hold the button
down to lower the foot end of the mattress platform
into a chair position.
The Bio-Contour up button simultaneously raises
the backrest and thigh sections to provide upright
SDWLHQWSUR¿OLQJWKHUDLVHGWKLJKVHFWLRQSUHYHQWV
the patient sliding down the bed.
The Bio-Contour down button returns the mattress
SODWIRUPWRDÀDWSRVLWLRQ
To operate controls
Select a function button for example
raise / lower
mattress platform height
. The indicator light on
the selected function will turn green and the bed will
immediately begin to move accordingly.
Press the Function Lock button, then press the
Attendant Control Panel button(s) corresponding to
the function(s) to be locked or unlocked.
Green light indicates function selected.
Yellow light indicates function is locked out.
Function lockout
Auto-Chair
Bio-Contour
Raise and lower the backrest.
The backrest will pause when it reaches an angle
approximately 30º above the horizontal.
Backrest
If the backrest angle is greater than 45º, it will return to 45º
to prevent the patient from tipping forwards.
The Auto-Chair down button returns the mattress platform
WRDÀDWDQGOHYHOSRVLWLRQ
Raise and lower the mattress platform. When
lowered to 38cm (40cm on beds with 150mm
FDVWRUVDERYHWKHÀRRULWZLOOSDXVHWKHQFRQWLQXH
to lower until it reaches its minimum height.
Mattress platform height
The “lock” indicator above the function button
will light yellow when the function is locked
and go out when it is unlocked.
At minimum height, clearance underneath the bed
is reduced. Keep your feet away from the areas
below the side rails and take extra care when
using patient hoists or similar equipment.
A “beeping” sound indicates
that the bed is operating on
battery power. Connect the bed
to the mains supply as soon as
possible.
Page 2/4