background image

Transport mode (mattresses only)

2

1

Activate:

Push the button down and pull the tubeset connector.

 

The mattress supports the patient for up to 12 hours.

Resume normal operation:

Reconnect the tubeset.

CPR control (mattresses only)

2

1

Activate:

 Press the two CPR release buttons 

simultaneously.

Reset: 

Push the front of the CPR unit until it clicks 

into place.

“Click”

Cleaning and disinfection 

Clean and disinfect between patients and at regular 
intervals while in use.

WARNING 

Remove the electrical supply to the pump before 
cleaning by disconnecting the mains power cord 
from the mains power supply or removing the 
battery pack.
Protective clothing should always be worn when 
carrying out decontamination.

Cleaning

1.  Clean all exposed surfaces and remove any debris by 

wiping with a cloth moistened with a neutral detergent 

and water. 

2.  Dry thoroughly. Do not allow water or cleaning 

solutions to collect on the surface of the pump.

Chemical disinfection

Use a chlorine releasing agent, such as sodium 
hypochlorite, at a strength of 1000 ppm available chlorine 
(from 250 ppm to 10 000 ppm depending on local policy 
and contamination status). 
Alcohol based disinfectants (strength 70%) may be used 

as an alternative. 

If an alternative disinfectant is selected, confirm suitability 
for use with the chemical supplier prior to use.
1.  Wipe all cleaned surfaces with the solution.
2.  Wipe with a cloth moistened in water.

3. 

Dry thoroughly.

Thermal disinfection

For laundering of the mattress top cover, see label.

Autofirm mode

Activate:

 Press and hold 

the button.

Comfort indicators flash 

and then illuminates in 

sequence.

Extend:

Comfort indicators

flash + series of tones: 

Press either comfort key 

to extend 5 minutes.

Maximum 30 minutes.

1

2

Deactivate:

 Press and 

hold the button. 

 2s

Pressure reached:

• 

Comfort indicators 

illuminated

• 

Pump beeps 3 times.

• 

Reactive (CLP) indicator 

illuminated

 2s

ArjoHuntleigh AB

Hans Michelsensgatan 10 

211 20 Malmö, Sweden

www.arjo.com

630937EN_01 • 06/2020 • Page 2/2

Reviews: