22
Measure
Each year/
10,000 cycles
Every other
year/20,000
cycles
Time*
(minutes)
2.7
&KHFNWKDWWKHRYHUÀRZGUDLQFRQQHFWLRQWRWKH
chamber is tight.
•
-
2.8
&OHDQWKHLQVLGHRIWKHÀXVKLQJFKDPEHU
•
10
2.9 Check the door and door spring.
•
15
3
Process tank
3.1 Clean as required.
•
-
3.2 Check and clean the level sensor.
•
5
3.3 Check that the valves are working and that the
connections are not leaking. Clean as required.
•
10
4
Steam Generators
4.1 Check the connections to the steam generator
for leakage and make sure the surrounding
insulation is intact and no hot surfaces are
exposed.
•
5
4.2 Check that the steam generator is working
properly.
•
10
5
Main pump
5.1
&KHFNWKDWWKHPDLQSXPSLV¿UPO\DWWDFKHG
and that there are no leaks in the connections.
•
5
6
Dosing system
6.1 Check the hoses between the dosing pump
and the detergent holder. Replace if necessary.
•
5–10
6.2 Check the dosage amount. Adjust if necessary.
•
10–30
6.3 Replace hose in the dosing pump.
•
5
6.4 Check that the empty container alarm works.
Clean or replace as necessary.
•
5
6.5 Check that the door to the detergent cabinet
closes properly (option IPX4).
•
1
6.6 Check the function of the check valve in the
steam generator.
•
5
7
Drainage connection
7.1
&KHFNWKDWWKHGUDLQDJHFRQQHFWLRQLV¿UPO\
attached and that it is not leaking.
•
5
8
Ventilator
8.1
(3$¿OWHU
•
10
8.2 Check the fan for proper operation and check
that the connections are tight.
•
5
8.3 Check that the check valves are functioning
properly. Clean as required.
•
10
8.4 Check hoses to the fan.
•
5
9
Automatic door
9.1 Check the automatic closing function (option).
•
5