Digital Scale

Model 1000 Scale

The 

MODEL 1000

 weight module is a compact and 

precise scale system integrated into the patient lifts 

and specifically designed for patient lifts

Completely self-contained, the MODEL 1000 scale 
combines the latest precision strain gauge technology 
and microprocessor signal conditioning to provide 
stable, accurate and repeatable weight data, with a 
calibrated accuracy of 0.1%. One-button Auto Zero 
Electronics makes resident/patient weighing quick 
and easy. The MODEL 1000 scale derives its power 
from a single 9-volts Lithium battery.

When battery replacement is required, an indicator 
will show on the display. Simply remove the four 
screws on the bottom of the case and remove the top 
casing to access the battery.

CAUTION:
The Load Cell contains no user serviceable
components and should be serviced by au-
thorized personnel only. Any unauthorized
tampering will void the warranty.

Digital Scale Calibration

1

The serial numbers for display and load cell must 
be identical.

2

Remove the screws from the back of the display 
to expose the electronics.

3

Adjust switch SW4 to the 

CAL.

 position 

(DOWN).

4

When no weight is applied to the system (includ-
ing sling), press SW1 (weight button) to adjust 
R3 (offset potentiometer) so that the display 
reads a +30 pound offset. Note: display will pe-
riodically flash 

CAL

. to let you know you are in 

CAL.

 mode.

5

Attach the sling to the load cell.

6

Press SW2 (zero button) to zero the display. 

7

Place a calibrated (must be at least Class F NIST 
traceable) weight on to the sling and adjust R10 
to that weight. Remove weight, reset zero (switch 
SW2) and repeat. It is recommended that you use 
at least 200 lb of calibrated weight to achieve an 
accurate calibration. 

8

When the reading of the display repeats the cor-
rect weight, remove all weight from the scale, 
then return switch SW4 to the run position (UP).

9

Press zero button (SW2).

10 When the system indicates “0.0”, place a cali-

brated weight onto the sling and make sure that 
the scale reads the correct value. Check scale at 
maximum weight. All readings must be within 
0.1% of NIST traceable standard.

SW4

SW3

SW1

SW2

R3

R10

CAUTION: 
Calibration shall be
performed by a
qualified technician
in order not to inva-
lidate the calibration
certificate. 

Summary of Contents for KM560001

Page 1: ...MAXI 500 OPERATING AND PRODUCT CARE INSTRUCTIONS 001 20815 EN rev 1 March 2006 Models KM560001 KM560101 KM560081 KM560181...

Page 2: ...available by contacting ARJO Hospital Equipment AB Our policy is one of continous development and we therefore reserve the right to make technical alter ations without notice The content of this publ...

Page 3: ...ansferring a patient 10 Battery pack 10 Removable battery pack 10 Emergency lowering 10 Battery charging 11 Battery and battery charger safety practices 12 Lifting a Resident Patient 13 Lifting a resi...

Page 4: ...l Equipment AB Box 61 S 241 21 Esl v Sweden Service and support A service routine has to be performed on your Maxi 500 by ARJO trained personnel to ensure the safety and operation of your product See...

Page 5: ...ibed in this manual The Maxi 500 shall always be handled by a trained caregiver who shall continuously attend to the resident patient during lift application and operation and in accordance with the i...

Page 6: ...ptional 3 Boom 4 Actuator 5 Emergency lowering 6 Safety latch 7 Two point spreader bar optional 8 DPS spreader bar optional 9 Right leg 10 Left leg 11 Mast 12 Battery pack 13 Charger 14 Start button 1...

Page 7: ...rofiles that can be used with the Maxi 500 Four point toileting sling with headrest Clip slings Four point padded sling Four point unpadded sling Four point mesh sling For more information about clip...

Page 8: ...les that can be used with the Maxi 500 Loop slings Hammock THA X For more information about loop slings and how to use them refer to the BHM Slings User Manual Hammock 6 THA6 X Quick fit TIR X Hygieni...

Page 9: ...slow walking pace 3 km hour or 0 8 meter second An assessment of the suitability for transfer of a patient who is connected to electrodes catheters or other medi cal devices must be conducted by qual...

Page 10: ...t to use the lift in an area that has a different voltage and cycle requirement Do not attempt to expose service or repair the lift battery or charger If any unit is malfunction ing contact your local...

Page 11: ...its functions and controls CAUTION Always carry out the items marked as Be fore each use in the Preventive Mainte nance Schedule before using the lift Power base models The base of the lift can be set...

Page 12: ...e hand control can be used to operate the lift To raise the patient Touch the UP button on the hand control Keep your finger on the button until the patient is at the desired height then release finge...

Page 13: ...volt 4 AH 24 volts capacity batteries delivering up to 35 lifts per charge 500 lb or 227 kg Battery life is variable 2 3 years and is influenced by proper charging practices and load exertion WARNING...

Page 14: ...owever if a partially charged battery is inserted the previous indicator level will be maintained even though the recently inserted battery may be in a better state of charge than indicated To achieve...

Page 15: ...dismantle or otherwise mechanically interfere with the batteries The power socket must be easily accessible If a fault condition occurs switch off the power cord and remove the connection plug from t...

Page 16: ...raise the resident patient using the hand control 7 The head support area of the sling should now come into use This is the most comfortable posi tion for transfer or transportation as it reduces pres...

Page 17: ...ing a resident patient from the floor Proper body mechanics and sound ergonomic positioning should be maintained by the caregiver at all times Clients being lifted from the floor are normally in this...

Page 18: ...15 METHOD 1 METHOD 2 METHOD 3 Lifting a resident patient...

Page 19: ...rews from the back of the display to expose the electronics 3 Adjust switch SW4 to the CAL position DOWN 4 When no weight is applied to the system includ ing sling press SW1 weight button to adjust R3...

Page 20: ...17 Parts List General Exploded View...

Page 21: ...20326 MAXI 500 RIGHT LEG COVER 1 15 000 20331 MAXI 500 LEFT LEG COVER 1 16 000 20501 BOLT CAP MAXI 500 2 17 001 16070 33 STICKER ARJO 1 18 001 20805 STICKER MAXI 500 2 19 001 20807 STICKER TOP BOOM MA...

Page 22: ...NC 2 9 000 05110 FLAT WASHER 0 0406 X 0 75 X 0 06 4 10 000 20065 SCREW M10 X 25 FLAT ZINC 2 11 000 20120 TIE ROD END KIT 213MM LR 1 12 000 20430 SWIVEL WHEEL 100MM 2 13 000 20440 SWIVEL WHEEL 100MM BR...

Page 23: ...20 Parts List Boom Exploded View Item Part Description eng Qty 1 000 20515 BOOM CAP MAXI 500 1 2 200 20252 BOOM BUSHING 2 3 236 20220 BOOM ASSY 1...

Page 24: ...X 7 Press the emergency stop button and make sure that all electrical power is cut off and that the green power light is off X X 8 Check all functions of the hand control ensure that the hand control...

Page 25: ...X 14 Check both mast bolts to ensure they are tight These bolts must be secured with lockwashers X 15 Check if the legs pivot bolts are secured with locknuts tighten if necessary to stabilize leg tra...

Page 26: ...ot to exceed the lifting capacity Charger Trouble Resolution Power on light on charger is not lit Check if the charger is plugged into the wall receptacle Charger is plugged in but Power on light is n...

Page 27: ...g Load S W L of 227 kg 500 lb 2 Product name 3 Read operating instructions before using the product 4 ARJO logo 5 Emergency stop and reset button identification 6 Date of manufacture product code seri...

Page 28: ...Control voltage output Vdc 24 Battery charger input Vac 100 to 240 Up and Down current limiting Amp 12 1 Duty cycle 10 6 min hour Sound power level dBA 40 DIGITAL SCALE SPECS Model 1000 Weight range 1...

Page 29: ...26 Technical Specifications...

Page 30: ...th Gateway Road Unit C Mississauga ONTARIO L4W 5J1 Tel 1 800 665 4831 Fax 1 800 309 1116 CESK REPUBLIKA ARJO Hospital Equipment s r o Strm 35 CZ 616 00 BRNO Tel 549 254 252 Fax 541 213 550 V If your c...

Reviews: