28
Instructions For Use
EN
It takes about 3.5 hours to charge the battery. It is recommended to place the
transducer in the docking slot when the transducer is not used for a long time.
Install the transducer in the base station and the transducer icon will be displayed on
the screen.
NOTE. A complete charging process takes approximately 3.5 hours.
2.3.3 General Application
Take out the transducer from the docking slot and it will power on automatically. The
transducer screen displays the signal strength, battery level and working channel.
After the transducer is successfully connected to the base station, it will also display
the transducer type. All the indicators are green. If the transducer is not successfully
connected, it will power off automatically.
Take the transducer up and keep the transducer at a distance of over 30cm from the
base station. The wireless connection indicator is on, and it indicates the transducer is
taken out. If you want to power off the transducer, put it back in the docking slot. If the
transducer connects to the base station successfully, the wireless connection indicator
is always on and do not put back the inactivated transducer in the docking slot. Place
the transducer on the patient
WARNING. The MECG cable is defibrillation-proof.
The DECG cable is not defibrillation-proof.
NOTE. Detailed operations please refer to the user manual of fetal&maternal
monitor.
NOTE. If the working status indicator is on, please do not put the uncharged
transducer in the docking slot.
NOTE. Fix the US-T transducer, TOCO-T transducer and TOCO transducer
tightly to ensure that they will not shift during movement.
NOTE. It is recommended that the transducer should be placed when the
patient stands for better monitoring
NOTE. Instruct the patient to move in the prescriptive area and distance for
obtaining better signal.
NOTE. The wireless ultrasound transducer may warm slightly (less than 3°C
(5.4ºF) above ambient temperature).
NOTE. The US-T transducer taken up first displays US1 on the screen, and
the one taken up later displays US2. Please do not take two US transducers
simultaneously and wait 2 seconds to take the other one. Restart the
transducers if you take up two US-T transducers at the same time by mistake.