40
Instructions For Use
EN
Appendix 1 Product Specifications
A1.1 Environmental Specifications
Operating
Storage
+5 ºC ~ +40 ºC
(+41ºF ~ +104ºF)
Temperature range
-20 ºC ~ +55 ºC
(-4ºF ~ +131 ºF)
15% to 93% (non
condensing)
Relative Humidity
15% to 93% (non
condensing)
86 kPa to 106 kPa
Pressure
70 kPa to 106 kPa
A1.2 Physical Specifications
Power Supply
Operating Voltage
100V-240V~
Operating Frequency
50Hz/60Hz
Input Power
0.8A-0.3A
Battery
14.8VDC/5000mAh
Standards Compliance
IEC 60601:2005+A1:2012, EN 60601-1:2006+A1:2013,
IEC 60601-1-2:2014, EN 60601-1-2:2015,
IEC/EN 60601-2-37, IEC 60601-2-27, EN 62479:2010,
ETSI EN 301 489-1, ETSI EN 301 489-3, ETSI EN 300
220-1, ETSI EN 300 220-2.
Anti-electric Shock Type
Class I equipment with internal power supply
Anti-electric Shock
Degree
FHR1, FHR2, TOCO BF
DECG
CF
MHR (from MECG)
CF with defib. protection
Degree of Protection
against
Harmful Ingress of Water
Base station: IPX1 (protected against vertically falling
water drops)
Transducers:
IPX8
continuous immersion in water 1.1m deep for 24 hours)
Degree of Safety in
Presence of Flammable
Gases
Equipment not suitable for use in presence of
flammable gases
Disinfection/Sterilizing
Method
Refer to this user manual for details
EMC
CISPR11 Group 1 Class A
Leakage Current
Ground Leakage
Current (Limit)
N.C. S.F.C.
500µA 1000µA