![Arjo Enterprise 9600X Instructions For Use Manual Download Page 45](http://html1.mh-extra.com/html/arjo/enterprise-9600x/enterprise-9600x_instructions-for-use-manual_2972338045.webp)
45
Initialisation
After connecting the bed to the electricity supply, the weighing
system must be initialised as follows:
1. Place the mattress, bed linen and all necessary
accessories on the bed. The patient must not be on the
bed at this point.
2. Press the
Zero
button once.
3. After a few seconds the display will read
.
4. The patient can now get onto the bed.
5. The bed cannot be re-zeroed once a patient is on the bed.
Use the Auto Compensation feature if adding or removing
items from the bed such as linen or accessories
WARNING
The weighing system is intended to provide reference data only. The
weighing system is not intended to provide readings upon which
medication dosage decisions are made.
Caution
The weighing system MUST be zeroed every time a new patient is to
be placed on the bed.
Caution
The weighing system MUST be zeroed every time a mattress is
changed.
Caution
The weighing system cannot be zeroed if a mattress or accessories
weighing more than 35kg (77 lb) are fitted. To ensure the bed can be
zeroed correctly, only use Arjo approved mattresses.