58
Technical Specifications
$IXOOWHFKQLFDOGHVFULSWLRQRIWKH
System 2000
FDQEHIRXQGLQWKHVHUYLFHGRFXPHQWDWLRQSOHDVHFRQWDFW$UMR
IRUDFRS\&RQWDFWLQIRUPDWLRQDSSHDUVRQWKHODVWSDJHRIWKLV,)8
Site Requirements
NOTE
All drainage, water and electrical installation work should be carried out by qualified personnel, in
accordance with local and national regulations.
Electric supply, Europe and North America
Voltage
230/240 V (Europe)
120 V (North America)
100 V (Japan)
220 V (Korea)
230 V (Saudi Arabia)
Number of phases
Single phase AC
Frequency
50 Hz (Europe)
60 Hz (North America)
50/60 Hz (Japan)
60 Hz (Korea)
60 Hz (Saudi Arabia)
Power consumption (maximum)
1100 VA
Fuse
6 A (Europe)
10 A (North America)
12 A (Japan)
6 A (Korea)
6 A (Saudi Arabia)
Protective earth
Yes
Potential equalisation
Yes
Power supply cord
Replaceable by qualified
personnel
WARNING
To avoid electric shock, make sure that the equipment is connected to:
• Continuously powered supply mains with protective earth.
• Separate fuse and ground fault circuit interrupter (GFCI)
• A mains disconnection device
• Equipotential bonding point
All installations must be in accordance with local codes and regulations.