Ardo medical AG

10. tECHniCal inFoRMation

10.1. Version

Carum with Euro mains cable 

Art. No. 63.00.71 Carum battery with Euro mains cable  Art. No. 63.00.75

Carum with UK mains cable 

Art. No. 63.00.72 Carum battery with UK mains cable 

Art. No. 63.00.76

Carum with UL mains cable 

Art. No. 63.00.73 Carum battery with UL mains cable 

Art. No. 63.00.77

10.2. technical data

Pump dimensions

L x B x H = 302 x 240 x 180 mm

Power cable length

Length 3 m

Pump weight

3 kg

Weight with battery

3.7 kg

Input voltage

100 – 230 VAC + / -10%

Frequency

50 – 60Hz

DC Input

12 VDC

Lifetime (excluding parts 
subject to wear and tear)

10 years

Operating 
temperatures

+5 – +40 °C

Relative humidity

10 – 93%

Protection level

BF

Air pressure

700 – 1060 hPa

Protection Category 
Double Insulated

Category II

Classification in 
accordance with 
MDD93 / 42

Category 2a

Cycle and vacuum levels

Vacuum range in 
stimulation mode

30 – 150 mbar  

22 – 98 mmHg 

 

  3 – 13 kPa

Cycle range in 
stimulation mode

72 – 120 / min.

Vacuum range in 
expression  mode 

30 – 330 mbar  

22 – 250 mmHg 

 

  3 – 33 kPa

Cycle range in expression  
mode

30 – 60 / min.

Permitted tolerance of 
the vacuum strength

+ / - 5%

Permitted tolerance of 
the cycles

+ / - 3 cycles

The vacuum level generated is sensitive to prevailing atmospheric pressure. The higher the prevailing 
atmospheric pressure, the more intense the generated vacuum – and vice versa. The vacuum values (including 
tolerances) specified in the user manual are based on an atmospheric pressure of 880hPa.

10.3. Compliance

In accordance with guidelines: MDD 93 / 42 CEE of 14.7.1993, and 2010, in respect of medical products 
and 2010. 2007 / 47 / CE, 2006 / 42 / CE, EN 60601-1, EN60601-1-2, EN ISO 10079-1, EN 60601-1-4,  
EN 60601-1-6, EN 60601-1-11

0123

10.4. EMV test records

See annex 12.1.

ARDO_GA_Carum_Inhalt_NEU.indd   27

12.10.12   10:48

Summary of Contents for Carum

Page 1: ... 42 54 Gebruiksaanwijzing Carum 55 67 Indicaciones Carum 68 80 Indicações Carum 81 93 Kullanım kılavuzu Carum 94 106 Brugsanvisning Carum 107 119 Bruksanvisning Carum 120 132 Bruksanvisning Carum 133 145 Käyttöohjeet Carum 146 158 Instrukcja Carum 159 171 Οδηγίες χρήσης Carum 172 184 Carum de en fr it nl es pt tr da sv no fi pl gr ...

Page 2: ...nnection opening for PumpSet c Control panel with LCD d Connection for mains cable e Bottle holder f EasyFreeze holder g Mains cable h Carum Container i Address field j Bottle holder k EasyFreeze holder l Carum m Mains cable n PumpSet f e g a c d j k l m n b h i ...

Page 3: ...the bottle holder 21 4 2 4 Installing the EasyFreeze holder 21 4 3 Operation 22 4 3 1 Switch on switch off 22 4 3 2 Regulation of the vacuum and display 22 4 3 3 Cycle regulation and display 22 4 3 4 Switch stimulation mode expression mode 23 4 3 5 Using express mode skipping stimulation mode 23 4 3 6 Timing Time Button 23 4 3 7 Backlight 23 5 USEFUL TIPS 24 5 1 Adjusting the vacuum intensity of s...

Page 4: ...a shower Keep the mains cable and all accessories away from sources of heat Do not use Carum in an area where oxygen gas is in use or sprays are being applied Only use a genuine ARDO PumpSet and genuine ARDO parts See Section 4 Start up Operation and Section 8 Maintenance Care Cleaning Carum should only be used for the purposes detailed in Section 2 Application Battery option and car battery adapt...

Page 5: ...istically stable over the entire vacuum and cycle area and corresponds closely to your baby s suckling behaviour Therefore the breast tissue is not additionally stressed by pumping Stimulation mode Exchange Stimulation Expression Expression mode 3 APPLICATION FOR USE The Carum breastpump system is intended for the expression of breast milk for the collection and storage of mother s milk in the eve...

Page 6: ... off Unit with the applied part BF Cleaning Single pumpset Double pumpset applied part Warning General risk points check the Instruction Manual Assembling 0123 Compliance with general guidelines MDD 93 42 EWG and MDD 2009 Manufacturer The unit is compliant with the stipulations of the WEEE 2002 96 EG guidelines Mains cable Device of protection category II IP21 IP 21 drip proof Li Ion Contains a li...

Page 7: ...unit D Fig 2 To use the double PumpSet Attach the tube connector to the unit Fig 3 The second opening of the tube connector E remains closed by single and double pumping 4 2 3 Installing the bottle holder The bottle holder is used for short term storage of the PumpSets The bottle holder is designed to protect against accidental spilling of milk The bottle holder can be used free standing or be plu...

Page 8: ...d off 4 3 2 Regulation of the vacuum and display The vacuum suction strength can be adjusted one level at a time by repeatedly pressing or by holding down the vacuum button to the left side of the key pad The vacuum strength can be individually regulated in both stimulation mode and expression mode Pressing the up arrow increases the vacuum pressing the down arrow reduces the vacuum The vacuum str...

Page 9: ...ycles from the previously set values When changing from express mode to stimulation mode the vacuum strength will be set to 60mbar and the cycles level to 90 cycles These values also apply when the unit is switched on 4 3 5 Using express mode skipping stimulation mode Sensitive Programme If the mode button is pressed within 2 seconds of switching on the Carum stimulation mode is skipped Express mo...

Page 10: ... milk ejection reflex For this reason the Carum will automatically start in stimulation mode i e with a higher cycles level and a lower vacuum strength mimicking the baby s natural behaviour Select stimulation mode choose the level of vacuum which feels most comfortable When the milk begins to flow the baby s sucking pattern changes The baby sucks more slowly but more strongly Change to express mo...

Page 11: ...ors when using the unit Weak or absent vacuum Disconnect the PumpSet from the unit Partly close the connector opening with a finger to test the suction If suction is de tectable then the pump is working Test the PumpSet in accordance with the separate instructions only use the original ARDO PumpSet The pump is not working no data on the LCD display Check the mains supply if the unit still does not...

Page 12: ...p for cleanliness After every renting Visual inspection of the pump for completeness After every renting Visual inspection of the pump for damage After every renting Simple operational checks of the pump Pump runs properly battery charged After every renting Check of the vacuum strength to be carried out by trained service personnel Minimum every 2 years 8 2 2 Checklist Maintenance inspection for ...

Page 13: ...ry II Classification in accordance with MDD93 42 Category 2a Cycle and vacuum levels Vacuum range in stimulation mode 30 150 mbar 22 98 mmHg 3 13 kPa Cycle range in stimulation mode 72 120 min Vacuum range in expression mode 30 330 mbar 22 250 mmHg 3 33 kPa Cycle range in expression mode 30 60 min Permitted tolerance of the vacuum strength 5 Permitted tolerance of the cycles 3 cycles The vacuum le...

Page 14: ...tions Carum und Carum with optional battery Temperature 20 70 C Relative humidity 10 80 Air pressure 700 1060 hPa 11 2 Disposal The unit may be returned to the manufacturer or be dismantled and recycled as metal or plastic in an environmentally friendly manner through a recycling firm specialising in electrical goods There are no special or increased risk factors Manufacturer Ardo medical AG Gewer...