Parameter
Setting
Alarm 2 Set
100 operations
With these settings, Alarm 1 is issued when the cumulative interruption counter for any of the three
phases dips below the set 1000 remaining operations ("Alarm 1 Set"). Similarly, when any of the
counters dips below 100 remaining operations, Alarm 2 is issued.
Events and registers
The circuit breaker wear function (abbreviated "CBW" in event block names) generates events and
registers from the status changes in Triggered, Alarm 1 and Alarm 2 signals as well as in internal pick-
up comparators. The user can select which event messages are stored in the main event buffer: ON,
OFF, or both.
The events triggered by the function are recorded with a time stamp and with process data values.
Table. 5.5.3 - 105. Event codes.
Event number
Event channel
Event block name
Event code
Description
3712
58
CBW1
0
CBWEAR1 Triggered
3713
58
CBW1
1
CBWEAR1 Alarm 1 ON
3714
58
CBW1
2
CBWEAR1 Alarm 1 OFF
3715
58
CBW1
3
CBWEAR1 Alarm 2 ON
3716
58
CBW1
4
CBWEAR1 Alarm 2 OFF
The function registers its operation into the last twelve (12) time-stamped registers. The register of the
function records the ON event process data. The table below presents the structure of the function's
register content.
Table. 5.5.3 - 106. Register content.
Date and
time
Event code
Trigger current
All.Op.ITrg
Deduct. Op
Op.Left
dd.mm.yyyy
hh:mm:ss.mss
3712-3716 Descr. Phase currents
on trigger time
Allowed operations
with trigger current
Deducted operations from
the cumulative sum
Operations
left
5.5.4 Fault locator (21FL)
The fault locator function is used for recording an estimated distance to the point where a fault has
occurred. It is mostly used in directional overcurrent protection or distance protection applications but
can be also triggered by other protections. The function can be used if all three phase currents and
three phase voltages have been connected to the relay. The triggering signals, the triggering current
and "Reactance per km" must be set in the configuration.
The inputs for the function are the following:
• operating mode selections
• setting parameters
• digital inputs and logic signals
• measured and pre-processed current magnitudes.
The function generates general time-stamped ON/OFF events to the common event buffer from each
of the two (2) output signals. The time stamp resolution is 1 ms. The function also provides a resettable
cumulative counter for the fault locator triggering events.
A
AQ
Q-S215
-S215
Instruction manual
Version: 2.04
202
Summary of Contents for AQ-S215
Page 1: ...AQ S215 Bay control IED Instruction manual...
Page 249: ...Figure 7 4 151 Example block scheme A AQ Q S215 S215 Instruction manual Version 2 04 248...
Page 269: ...Figure 8 13 172 Device installation A AQ Q S215 S215 Instruction manual Version 2 04 268...
Page 287: ...10 Ordering information A AQ Q S215 S215 Instruction manual Version 2 04 286...