instantaneous voltage. The minimum trigger voltage, as a function
of frequency, is shown in the following graph:
NOTE:
The minimum rms voltage at 60 Hz is (approx.) 11 volts.
“
AC Removed
” configures the timer to trigger on the removal of
an ac voltage, as described above. The point in time at which the
signal last falls below the threshold is interpreted as the trigger.
Slight errors may occur, when using this mode of triggering, that
depend on the instantaneous trigger voltage. However, this error
should not exceed 1 millisecond.
“
Contact Open
” configures the timer input to recognize the
opening of an electrical contact as a trigger event. The threshold
is (approx.) 2.9 volts, with a pull-up impedance of (approx.) 47
kilohm to a source voltage of (approx.) 8 volts.
“
Contact Closed
configures the timer input to recognize the closing
of an electrical contact as a trigger event. The threshold is (approx.)
2.6 volts, with a pull-up impedance of (approx.) 47 kilohm to a
source voltage of (approx.) 8 volts.
3.7.2.5
Select Timer Trigger B
Press the Timer CONFIG key and select Trigger B from the Timer
Configuration Menu and press ENTER. Select the appropriate trigger
signal type for timer “Channel B” from this menu. The choices for input
“B” are exactly the same as those for input “A”. See Section 3.7.2.4.
3.7.3
Timer Measurement Categories
All measurements performed by the timer section of the Model 931A can be
classified as either
time
or
count
measurements. The following paragraphs
describe both categories.
62
Summary of Contents for 931A
Page 6: ...vi ...
Page 12: ...xii ...
Page 14: ...xiv ...
Page 16: ...xvi ...
Page 52: ...SOURCE LOAD B A C N Figure 3 12 3 Phase 4 Wire 3 Element Connections 36 ...
Page 90: ...Iout I1 V1 V2 V3 I2 I3 Figure 3 17 3 Phase 4 Wire 3 Element Transducer Test 74 ...
Page 114: ...98 ...
Page 120: ...104 ...
Page 142: ...126 ...