A.3.32
Range Hold, Set (RHD)
Command:
ch#RHD
Freezes the current analog input range setting for the selected channel.
Ch#
To select
1
Channel 1
2
Channel 2
A.3.33
Logged Data, Review (RLD)
Command:
#RLD
Selects a specific stored data record for display if # is included in the
command, otherwise returns a list of the stored data records by #, date,
time and type. Store data records manually using the log data key, or via
RS-232C using the LOG command. Use 0RLD to clear all logged data.
When issuing this command, the specified record is written to the
Model 931A display, overwriting current measurement data and placing
the Model 931A in Measurement Hold Mode.
Use 0MHD to release
Measurement Hold Mode, and resume measurement.
See appendix B for use of this command with the Extended Memory
Option.
A.3.34
Range Hold, Release (RRH)
Command:
ch#RRH
Releases range hold for the selected channel. Places the selected channel
in auto range as necessary.
ch#
select
1
Channel 1
2
Channel 2
A.3.35
MIN/MAX, Restart (RST)
Command:
RST
Restarts the minimum and maximum accumulation. Will also ”zero” the
Energy Readings.