Special Functions and Features
39
Voltage:
Display Digits
Range
Low
High
1 2 3 4 5 6 7 8
V 1
0.2V
2V
±
X . X X X
O
L
2
2V
20V
(
<
10V)
±
X . X X X
T
(
≥
10V)
±
X X . X X
S
3
20V
200V
(
<
100V)
±
X X . X X
(
≥
100V)
±
X X X . X
4
200V
2000V
*
(
<
1000V)
±
X X X . X
(
≥
1000V)
±
X X X X
*Maximum specified input voltage is 1000V, rms or dc.
3.1.2 Automatic Range Selection
When the instrument is first turned on, it will default to automatic
range selection. The operating ranges are selected automatically,
depending on the levels of voltage and current present on the inputs.
During the transition from one range to the next, the display will
momentarily blank, indicating that a range change is taking place.
3.1.3 Manual Range Selection
The 918B has a provision for manually selecting and holding a
particular measurement range. This can be done for either the
voltage or current mode, or both.
Procedure
1.
To select and hold a range, first switch the instrument into the
desired mode (i.e. voltage or current).