122
Functional Description
11.28.1
Signal Functions Available for Triggering
ID No.
Function
Update Rate
Channels Available
0
Off
N/A
N/A
1
Voltage
20/sec
A, C, B, Average
2
Current
20/sec
A, C, B, Average
3
Active Power (Watts)
20/sec
A, C, B, Total
4
Reactive Power (VARs)
20/sec
A, C, B, Total
c
5
Apparent Power (VA)
20/sec
A, C, B, Total
c
6
Power Factor (PF)
20/sec
A, C, B, Total
c
7
Sequence Voltage Components
20/sec
Zero, Pos., Neg.
8
Sequence Current Components
20/sec
Zero, Pos., Neg.
9
Phase Balance Ratio, derived from
sequence components
20/sec
Zero/pos., Neg./Pos.
for both voltage & Current
10
Frequency and Time
20/sec
Frequency, F, df/dt, T
11
THD, Voltage
1/sec
A, C, B, max (A, C, B)
a
12
THD, Current
1/sec
A, C, B, max (A, C, B)
b
13
Harmonic VRMS
1/sec
A, C, B, max (A, C, B)
a
14
Harmonic IRMS
1/sec
A, C, B, max (A, C, B)
b
15
K-Factor, voltage
1/sec
A, C, B, max (A, C, B)
a
16
K-Factor, current
1/sec
A, C, B, max (A, C, B)
b
17
Flicker, instantaneous, voltage
1/sec
A, C, B, max (A, C, B)
a
18
Flicker, PST, voltage
1/sec
A, C, B, max (A, C, B)
a
19
Flicker, instantaneous, current
1/sec
A, C, B, max (A, C, B)
b
20
Flicker, PST, current
1/sec
A, C, B, max (A, C, B)
b
21-25
Reserved
TBD
TBD
26-31
Transient detector channels 0 - 5
20/sec
Bound, rate/change, osc.
Table 11.2: Signal Functions Available for Triggering
a
Average or maximum of 3 voltage channels in 3Φ4W3E and 3Φ3W2E input mode, 1 channel in 1Φ2W1E mode,
and 2 channels in other modes. In 3Φ3W2E mode, the voltage VAC is derived internally by point-by-point calculation
after A/D conversion. The voltage VB, synthesized internally in 3Φ4W2-1/2E mode, does not correspond to an actual
physical quantity and is not included in the analysis, although measurements can be made on the synthesized signal.
b
Average or maximum of 3 current channels in 3Φ modes, 1 channel in 1Φ2W3E mode, and 2 channels in other
modes. In 3Φ3W2E mode, the current IB is derived internally by point-by-point calculation after A/D conversion.
c
Total of 3 elements in 3Φ4W3E and 3Φ4W2-1/2E modes, 1 element in 1Φ2W1E mode, and 2 elements in other
modes.