Chapter 3
Physical Configuration
3.1
Instrument
3.1.1
Location Considerations
The Model 1073A is designed to operate in an environment having an ambient temperature range
of 0
◦
C to +50
◦
C (+32
◦
F to +122
◦
F). Operation is possible at temperatures of
−
20
◦
C to +70
◦
C,
typical. No external ventilation is necessary.
For best access to the 1073A, allow adequate clearance for rear-panel connections, especially in
rack-mounting situations.
3.1.2
Power Requirements
The standard AC input voltage range for the Model 1073A is 85 Vac to 264 Vac, 47 Hz to 440 Hz,
110 to 275 Vdc, or 10 to 60 Vdc. See Section 3.1.3.
3.1.3
Power Line Connection
The Model 1073A is equipped with a user specified Power Input Module. The modules available
are described in the following paragraphs.
Option 07, AC/DC Power Input Module (IEC-320 Inlet)
AC Operation
This power inlet module operates across an AC input voltage range of 85 – 264 Vac, 47 – 440 Hz.
The mating cordset provided is dependent upon the cordset option (P1 through P10) which was
specified at the time of purchase. For further information about available cord sets for specific
countries, see Table 1.1, or contact Arbiter Systems.
To connect the input power, first plug the end of the power cord having the mating IEC
connector into the power inlet module on the rear panel, then plug the other end into an appropriate
power outlet.
WARNING: For maximum safety and best performance, always connect the
input cord to a properly grounded power source.