background image

4

Phase Angle

Calibration of phase angle meters or

transducers for single phase operation
requires use of both the voltage and current
outputs of the 1040C.  The representative
circuit connection diagrams for these types of
devices are the same as for the ac power
meters and transducers, given in figure 5.
Typical connections for calibration are also
the same as for ac power meters and
transducers, and are given in figure 6.  For
calibration, the 1040C voltage output leads
are connected to the voltage terminals of the
meter or transducer, and the 1040C current
output leads are connected to the current
input terminals of the meter or transducer
(figure 6a and 6c).  If the meter or transducer
uses an external current transformer, one of
the current output leads of the 1040C should
be passed through the center of the current
transformer, and then shorted to the other
lead.  If the meter or transducer uses an
external potential transformer, then likewise
it must be inserted between the 1040C
voltage output leads and the input voltage
leads of the unit under test. Refer to figures
6b and 6d for these two types of connections.
The 1040C should first be configured to
operate in the ac power mode, as described in
paragraph 2.5.6, and the appropriate voltage
and current values should be selected.  After
the outputs have been enabled and the phase
angle meter or transducer under test has been
observed to be indicating zero, press the
"Phase" button.  This will allow shifting of
the relative phase between the voltage and
the current outputs.  The desired phase angle
can now be entered directly, via the keypad,
and followed by the appropriate units
(degrees).  A negative value for phase angle
corresponds to the output current lagging the
output voltage.  After entering the value for
phase angle, the operator may adjust the
value by using the control knob, or by using
the "Up" or "Down" keys.  For more
complete instructions on operation of the

1040C for phase angle calibrations, refer to
section 2.5.1 of the operation manual.

Power Factor

The Power Factor in a circuit is equal to

the cosine of the phase angle between the
voltage and the current. With this in mind,
any power factor can be simulated by
adjusting the phase relationship between the
voltage and current, which can be
accomplished quite easily with the 1040C.
Two methods can be used:

 

The required phase angle can be entered
directly via the keypad, as described in
the section on phase angle calibration;

 

The power factor can be entered directly,
by first pressing the "Power Factor" key,
then using the adjust knob to vary the
displayed power factor.  This method
allows entry of lead or lag power factor
values without having to first calculate
the corresponding phase angle.

For an illustration of the relationship

between phase angle and power factor in the
1040C, refer to figure 12.

In order to accomplish power factor

indication, some transducer manufacturers
recommend using one of their phase angle
transducers and performing a mathematical
conversion on the output signal.  This can be
done manually, using conversion tables
supplied by the manufacturer, or
automatically, using a meter having cosine
scaling.  Typical transducer application
circuit connections and calibration setup
connections are usually the same as those for
a phase angle, watt, or var transducer, as
shown in figures 5 and 6.  Calibration can be
performed using the connections in figure 6
and one of the methods outlined above, and
in accordance with the transducer
manufacturers recommendations.

Power factor output is also commonly

available as an additional feature on single

Summary of Contents for 1040C

Page 1: ...0800A 9 92 CALIBRATING METERS AND TRANSDUCERS WITH THE MODEL 1040C PANEL METER CALIBRATOR ARBITER SYSTEMS INC 1324 VENDELS CIRCLE SUITE 121 PASO ROBLES CA 93446 805 237 3831 FAX 805 238 5717 PD0010800A 9 92 ...

Page 2: ...r 4 Volt Amps and Volt Amps Reactive 5 Using the 1040C for Calibrating Polyphase Devices 7 A C Voltage 7 A C Current 7 A C Power 7 Phase Angle 8 Power Factor 8 Volt Amps and Volt Amps Reactive 10 Calibrating Transducers Using the Null Comparison Method 11 Connection Diagrams 12 Phase Power Factor and VAR Convention Diagram 23 Appendix A Meter Examples 24 ...

Page 3: ...ion and distribution The 1040C Panel Meter Calibrator can be configured to perform eight calibration functions Voltage ac and dc Current ac and dc Frequency Power Power Factor Phase VARs Synchroscope A front panel keyboard and display allow direct selection of values for all of the parameters shown above Entered values can be further modified using a knob also located on the front panel A percent ...

Page 4: ...operation manual A C Current Calibration of a single phase transducer or meter for ac current follows virtually the same process as for ac voltage meters and transducers except that only the current output of the 1040C is used rather than only the voltage output Representative schematics are given in figure 3 for typical circuit installations of single phase ac current meters and transducers Conne...

Page 5: ...ed by A After this is completed pressing the Operate key will enable the outputs which will increase to the respective values entered An equal sign on the front panel display indicates that the current and voltage outputs and the phase angle between them have stabilized at the proper values The calibration procedure for the meter or transducer under test can then be performed The power value can b...

Page 6: ...to the output current lagging the output voltage After entering the value for phase angle the operator may adjust the value by using the control knob or by using the Up or Down keys For more complete instructions on operation of the 1040C for phase angle calibrations refer to section 2 5 1 of the operation manual Power Factor The Power Factor in a circuit is equal to the cosine of the phase angle ...

Page 7: ... Vars are calculated as the product of the RMS voltage the RMS current and the sine of the phase angle between the two For an illustration of the relationship between phase angle and vars when using the 1040C refer to figure 12 The 1040C has a provision for displaying and modifying both volt amps and vars successively pressing the Power key during operation will first display watts then vars then ...

Page 8: ...6 detailed instructions on operating the 1040C in the volt amps and volt amps reactive modes refer to section 2 5 6 of the operation manual ...

Page 9: ...so be connected to the 1040C in the same manner as shown in figure 8b with the same total voltage burden restriction A C Power Power measurement in a three phase four wire system requires three complete watt meters or watt transducers each watt meter or transducer consists of one current element and one voltage element Figure 9a shows a typical connection diagram for measurements of this type A cu...

Page 10: ...he transducer and the 1040C current output leads are connected to the current input terminals of the meter or transducer If the transducer uses an external current transformer one of the current output leads of the 1040C should be passed through the center of the current transformer and then shorted to the other lead If the meter or transducer uses an external potential transformer then likewise i...

Page 11: ...to vary the displayed power factor This method allows entry of lead or lag power factor values without having to first calculate the corresponding phase angle Calibration of the transducer should be performed according to the instrument manufacturer s recom mendations Typical applications of power factor meters for multi phase systems involve circuit connections which are similar if not identical ...

Page 12: ...ly be modified to a point within the current range of the 1040C If the necessary current exceeds this range the display will indicate Entry Error Additionally if the phase angle is set to zero and the user attempts to enter any var value other than zero Entry Error will be indicated This is because under these conditions no value of voltage or current will produce anything other than zero vars Typ...

Page 13: ...tial meter dislays zero The advantage of this method is that since the input signals are exactly the same the accuracy of the source and the equipment used to measure it are inconsequential The calibration accuracy is dependent only on the individual accuracies of the precision standard and of the differential meter Precision standards are usually available from the manufacturers of the transducer...

Page 14: ...ducer Typical Installation no Potential Transformer L O A D Figure 1 Single Phase Voltage Meters and Transducers Typical Circuit Connections V L I N E L1 N L1 N b Voltmeter Typical Installation with Potential Transformer L O A D V Inst Pwr Output L1 N L I N E L1 N d Voltage Transducer Typical Installation with Potential Transformer L O A D ...

Page 15: ...ions no Potential Transformer Figure 2 Single Phase Voltage Meters and Transducers Typical Circuit Connections 1040C Vout 1040C Vout R B V Inst Pwr Output d Voltage Transducer Calibration Connections with Potential Transformer 1040C Vout R B V b Voltmeter Calibration Connections with Potential Transformer 1040C Vout R B R B ...

Page 16: ...tput L1 N L O A D L1 N L I N E c Current Transducer Typical Installation no Current Transformer d Current Transducer Typical Installation with Current Transformer L1 N L1 N L I N E L O A D Figure 3 Single Phase Current Meters and Transducers Typical Circuit Connections L I N E N L1 L1 N L O A D Inst Pwr Output L I N E N L1 L1 N L O A D ...

Page 17: ...nt Transformer c Current Transducer Calibration Connections no Current Transformer d Current Transducer Calibration Connections with Current Transformer Figure 4 Single Phase Current Meters and Transducers Calibration Connections R B 1040C out 1040C out R B Inst Pwr Output 1040C out R B Inst Pwr Output 1040C out R B ...

Page 18: ...ucer Typical Installation with Current and Potential Transformers Figure 5 Single Phase Watt Var or Power Factor Meters and Watt Var Power Factor or Phase Transducers Typical Circuit Connections Current Connections Voltage Connections Current Connections Voltage Connections L1 N L I N E L O A D L1 N W Var PF Current Connections Voltage Connections L1 L1 N N L O A D L I N E L1 N L I N E L O A D L1 ...

Page 19: ...nt or Potential Transformers d W Var PF PA Transducer Calibration Connections with Current and Potential Transformers Figure 6 Single Phase Watt Var or Power Factor Meters and Watt Var Power Factor or Phase Transducers Calibration Connections R B R B 1040C out Vout Current Connections Voltage Connections Current Connections Voltage Connections R B B R 1040C out Vout 1040C out Vout R B R B Current ...

Page 20: ... 1 1 1 1 ZA ZB ZC Total Burden Current Where V Output Voltage of 1040C Zx Impedance of Transducer or Meter x at Frequency of Operation c Burden Calculation b Meters 1040C out Vout VA VB VC R B 1040C out Vout A B C Vin Vin Vin Outputs To DVM s R B a Transducers Inst Pwr ...

Page 21: ...cers or Meters Total Burden Voltage Where Output Current of 1040C Zx Impedance of Transducer x at Frequency of Operation c Burden Calculations 1040C out VA VB VC b Meters a Transducers 1040C out A B C in in in Outputs To DVM s R B Inst Pwr ZA ZB ZC R B ...

Page 22: ... Power Measurement in a 3 Phase 4 Wire System using a 3 Element Transducer L1 L2 L I N E L1 L2 L O A D L3 N in Vin in in Vin Vin L3 N Inst Pwr Output L1 L2 L I N E L3 in Vin in Vin L1 L2 L O A D L3 A B b Circuit Connections for Power Measurement in a 3 Phase 3 Wire System Using a 2 Element Transducer ...

Page 23: ...1 Figure 10 Calibration Connections for 3 Element Watt Transducer Current and Potential Transformers are included for reference Inst Pwr Output to D V M A B C in Vin in in Vin Vin 1040C Vout out R B R B ...

Page 24: ... Typical Circuit Connections L1 L2 L I N E L1 L2 L O A D Inst Pwr Output to D V M in Vin L3 N L3 N Note Wiring Also Applies for 3 Phase 3 Wire System Inst Pwr Output to D V M in Vin 1040C Vout out b Typical Calibration Connections R B R B ...

Page 25: ... PF 1 0 Vars 0 W VA Phase 135 PF 707 Lag Vars 707 VA W 707 VA Phase 45 PF 707 Lag Vars 707 VA W 707 VA Conventions Used in the 1040C Power Factor Cos θ Vars Vrms Irms Sin θ Watts Vrms Irms Cos θ VA Vrms rms Where θ Theta Phase Angle between voltage and current A negative number indicates that current is lagging voltage Example Figure 12 Phase Power Factor and VAR Conventions Employed by the 1040C ...

Page 26: ...he following pages contain connection diagrams for use in calibrating various types of Yokogawa panel meters These illustrations are for reference only the manufacturer s data sheets should be consulted to insure proper connections ...

Page 27: ...25 AUX V VOLTAGE CURRENT 1040C PANEL METER CALIBRATOR REMOTE Black Red AC AMMETER YOKOGAWA 103131 AUX V VOLTAGE CURRENT 1040C PANEL METER CALIBRATOR REMOTE Black Red AC VOLTMETER YOKOGAWA 103021 ...

Page 28: ...0C PANEL METER CALIBRATOR REMOTE Red Black Black Red 3 WIRE 3 PHASE WATTMETER YOKOGAWA 10322 103712 1 4 5 6 9 11 7 8 2 3 4 WIRE 3 PHASE WATTMETER AUX V VOLTAGE CURRENT 1040C PANEL METER CALIBRATOR REMOTE Red Black Red Black YOKOGAWA 10325 103732 ...

Page 29: ...RENT 1040C PANEL METER CALIBRATOR REMOTE Red Black Black Red YOKOGAWA 10321 103702 10331 103762 WATT WATT VAR VAR 1 4 5 6 7 8 3 2 3 WIRE 3 PHASE VARMETER AUX V VOLTAGE CURRENT 1040C PANEL METER CALIBRATOR REMOTE Black Red Red Black YOKOGAWA 103812 10328 10332 103772 ...

Page 30: ...VOLTAGE CURRENT 1040C PANEL METER CALIBRATOR REMOTE Red Red Black Black 4 WIRE 3 PHASE 2 1 2 ELEMENT VARMETER YOKOGAWA 10329 103742 FREQUENCY METER AUX V VOLTAGE CURRENT 1040C PANEL METER CALIBRATOR REMOTE YOKOGAWA 103372 Black Red ...

Page 31: ...ETER CALIBRATOR REMOTE Red Black Black Red SINGLE PHASE POWER FACTOR METER YOKOGAWA 103412 1 4 5 6 7 8 3 2 3 PHASE POWER FACTOR METER FOR BALANCED SYSTEM AUX V VOLTAGE CURRENT 1040C PANEL METER CALIBRATOR REMOTE Red Red Black Black YOKOGAWA 103462 ...

Page 32: ...30 1 4 5 6 9 11 7 8 2 3 4 WIRE 3 PHASE POWER FACTOR METER AUX V VOLTAGE CURRENT 1040C PANEL METER CALIBRATOR REMOTE Black Red Black Red YOKOGAWA 103472 ...

Reviews: