AR DC6180 Operating And Service Manual Download Page 15

 

 

041910 

160 School House Road Souderton, PA 18964-9990 • 215-723-8181 • www.ar-worldwide.com 

 

Model DC6180A, M1 

Model DC6180, M1 through M6 

Dual Directional Coupler 

600 Watts CW 

80MHz–1000MHz 

 

The Model DC6180A is a dual directional coupler with excellent frequency range and power handling 
capability.  The wide range assures flexibility in coupling medium power amplifiers to oscilloscopes, voltmeters, 
power meters, spectrum analyzers and other measuring instruments.  The dual directional design allows the 
user to monitor both forward and reflected power, a much needed characteristic in RF susceptibility testing for 
amplifier overdrive protection, field control and load protection.  Low insertion loss allows efficient coupling to 

the load while the coupling factor is well matched to the typical laboratory equipment. 

 

The Model DC6180A is recommended for use with our high power W Series Amplifiers. 

 

SPECIFICATIONS 

 

FREQUENCY RANGE .............................................80–1000 MHz 

 

POWER (maximum) .................................................See Model Configuration 

 

COUPLING FACTOR ..............................................See Model Configuration 

 

COUPLING FLATNESS ............................................± 0.5dB 

 

DIRECTIVITY 
 Typical 

.............................................................25 

dB 

 Minimum 

.........................................................20 

dB 

 

INSERTION LOSS  ..................................................0.15 dB maximum 

 

IMPEDANCE (Main Line) .........................................50 ohms, VSWR 1.15:1 maximum 

 

CONNECTORS ......................................................See Model Configuration 

 

WEIGHT ................................................................0.6 kg, 1.2 lb 

 

SIZE (WxHxD) .........................................................10.9 x 6.3 x 3.2 cm (4.3 x 2.5 x 1.3 in) 

  

MODEL CONFIGURATION 

 

Model Number 

Mainline Connector 

J1/J2 

Coupled Connector 

J3/J4 

Power (Watts) 

CW/Peak 

Coupling Factor, 

including flatness (dB) 

DC6180A N(M)/N(F) 

N(F)/N(F) 

600/1000 

60±1.0 

DC6180AM1* N(M)/N(F)  N(F)/N(F) 600/1000  60±1.0 
DC6180 N(F)/N(F) 

BNC(F)/BNC(F) 

600/1000 

60±1.0 

DC6180M1 N(F)/N(F) 

N(F)/N(F) 

600/1000 60±1.0 

DC6180M2 N(F)/N(F) 

SMA(F)/SMA(F) 

600/1000 60±1.0 

DC6180M3 

See DC6180M3 Specification Sheet 

DC6180M4 

See DC6180M4 Specification Sheet 

DC6180M5 

See DC6180M5 Specification Sheet 

DC6180M6 N(F)/N(F) 

N(F)/N(F) 

500/1000 50±1.0 

*

DC6180AM1 includes A2LA calibration

 

Summary of Contents for DC6180

Page 1: ...Operating and Service Manual DC6180A Model 10011701 Part Number Serial Number 160 School House Road Souderton PA 18964 215 723 8181 Fax 866 859 0582 www arworld us...

Page 2: ......

Page 3: ...riction of the use of certain hazardous substances in electrical and electronic equipment recast RECYCLING WEEE DIRECTIVE 2012 19 EU OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of 4 July 2012 on was...

Page 4: ......

Page 5: ...clearance on all vented sides of the equipment to permit the airflow required for proper ventilation Do not restrict airflow into the equipment by blocking any vents or air intakes Restricting airflow...

Page 6: ...a assembly may be heavy requiring two persons to lift Use caution when installing or removing unit Follow all equipment setup and lifting instructions specified in this document Ensure connectors are...

Page 7: ...von 10 2 cm 4 in auf allen bel fteten Seiten des Ger ts ein um eine ausreichende Luftzirkulation zu gew hrleisten Beintr chtigen Sie den Luftstrom des Ger ts nicht indem Sie L ftungs ffnungen oder de...

Page 8: ...allen in diesem Dokument angegebenen Anweisungen zur Anbringung Stellen Sie sicher dass alle Anschl sse f r den beabsichtigten Betrieb geeignet sind Informationen zu den Anschl ssen erhalten Sie im Be...

Page 9: ...tilation Ne limitez pas le flux d air allant dans l quipement en bloquant tout vent ou entr e d air La restriction du flux d air peut endommager l quipement causer des coupures intermittentes ou des d...

Page 10: ...dent lorsque vous installez ou retirez l unit Respectez toutes les instructions concernant l installation et le levage de l quipement indiqu es dans ce document Assurez vous que les connecteurs sont a...

Page 11: ...is voor goede ventilatie mogelijk te maken Belemmer de luchtstroom in het apparaat niet door ventilatieopeningen of luchtinlaten te blokkeren Het belemmeren van de luchtstroom kan leiden tot schade a...

Page 12: ...nstellen en optillen van de apparatuur die in dit document worden vermeld Zorg ervoor dat de connectoren geschikt zijn voor de beoogde werking De connectoren worden gespecificeerd in de gebruikershand...

Page 13: ...is quite simple The input signal is fed into the N male connector J1 marked INPUT and the output is taken from the N female connector J2 marked OUTPUT The input and output jacks on the unit are Type...

Page 14: ...Model DC6180A 2 Rev B...

Page 15: ...The Model DC6180A is recommended for use with our high power W Series Amplifiers SPECIFICATIONS FREQUENCY RANGE 80 1000 MHz POWER maximum See Model Configuration COUPLING FACTOR See Model Configuratio...

Page 16: ......

Page 17: ...Model DC6180A Rev B 5...

Page 18: ...Model DC6180A 6 Rev B...

Page 19: ...cover If the security label is removed and or cut the warranty of the module will be voided Vacuum tubes in the L series amplifiers traveling wave tubes in TWT amplifiers and power heads will be free...

Page 20: ......

Reviews: