Model 125S1G2z5
Rev A
31
4.3.1
General - Reading Faults
The Model 125S1G2z5 incorporates relatively simple fault detection circuitry, which makes use of the digital
display panel to alert the user or technician which component(s) need service. Use of these indications can
usually expedite troubleshooting of the amplifier. Most faults can be immediately determined down to the
assembly level. If a reset is still indicated, turn off the RF power signal to the input of the amplifier, read and
record the fault indication displayed on the digital control panel for later reference. Then, use the RESET
function to see if the fault clears. If the fault clears, slowly bring the amplifier’s drive level back up and
ensure that recommended RF power levels are not exceeded. If the fault indication is no longer visible, the
fault may have been brought about by a temporary transient condition, component thermal condition or
excessive RF drive to the amplifier’s input. If the fault does not clear with the RF drive off, some other
problem exists in the amplifier.
4.3.2
The Unit Cannot Be Operated Remotely
1.
Verify that the front panel LOCAL/REMOTE switch is set to the REMOTE position.
2.
Verify that the unit operates locally by resetting the LOCAL/REMOTE switch to the LOCAL position; if
the unit does not operate locally, see 4.3.1 of this manual.
3.
Check the position of the ADDRESS via the control software on the front panel touch screen as shown in
Figure 2-3. Check to see that these switches are properly set for either RS-232 or IEEE-488 operation, as
desired. (See Section 2 of this manual for the proper ADDRESS switch settings.)
4.3.3
Thermal Fault (Schematics 10047173, 10033124)
During a Thermal Fault, the front panel touch screen interface should read
Ax THERMAL FAULT
.
1.
Try to reset the unit; if the unit resets and operates normally, check to see that the cooling fans (B1–B2)
are operating normally and that the air inlet on the bottom of the unit and the air outlets on the rear of the
unit are not blocked.
2.
If the unit does not reset and the cooling fans is operating normally, check the voltage at the A12
Control/Fault Board, J5, Pin 23, 26 or 27 should be
≤
0.1V.
3.
If the voltage on A12 J5, pin 23, 26 or 27 is high, check the connection to J3-5 and C53.
Summary of Contents for 10047590
Page 2: ......
Page 4: ......
Page 6: ......
Page 21: ...Page 3 125S1G2z5 125 watts CW 1 0GHz 2 5GHz Graphs...
Page 22: ...Page 4 125S1G2z5 125 watts CW 1 0GHz 2 5GHz Graphs...
Page 24: ......
Page 40: ...Model 125S1G2z5 24 Rev A...
Page 44: ...Model 125S1G2z5 28 Rev A...
Page 54: ......