FL7006/FL7030/FL7218/FL7040/FL7060Kit
24
Rev J
3.7.1
RS-232 Operating Protocols
3.7.1.1
Normal Operation Command/Response
This section lists the operating protocols RS-232 on the interface for normal operation.
Communication Protocol
Data Type:
RS-232 Serial
Data mode:
Asynchronous
Word Length:
8 bit
Parity:
None
Stop Bits:
1
Data Rate:
19,200 baud
RS-232 (DCE) Port Pinout Diagram DB-9 Female
Pin 1
DCD
Pin 2
TD
Pin 3
RD
Pin 4
DTR
Pin 5
GND
Pin 6
DSR
Pin 7
CTS
Pin 8
RTS
Pin 9
Unused
Information Transfer Protocol
The probe kit only responds to commands from another device. It transmits no data without first receiving
instructions to do so.
All characters are defined by and transferred as common 7 bit ASCII characters.
Command Structure
A command string consists, in sequence, of:
•
command letter
•
parameter (optional)
•
carriage return <CR>
Response Structure
In response to a command, the interface typically responds with a response string, in sequence, of:
•
start character colon <:>
•
command letter of the received command
•
data (optional)
•
termination character(s)
Note: Beginning with FI7000 firmware revision 2.70, the user can select the
termination character(s) used for response strings.
Summary of Contents for 10013164
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 20: ......
Page 24: ...FL7006 FL7030 FL7218 FL7040 FL7060Kit 12 Rev J ...
Page 44: ...FL7006 FL7030 FL7218 FL7040 FL7060Kit 32 Rev J ...
Page 48: ...FL7006 FL7030 FL7218 FL7040 FL7060Kit 36 Rev J ...
Page 50: ...FL7006 FL7030 FL7218 FL7040 FL7060Kit 38 Rev J ...
Page 51: ...Rev J 39 Appendix B Dimensional Drawings Figure B 1 Models FL7030 FL7006 ...
Page 52: ...FL7006 FL7030 FL7218 FL7040 FL7060Kit 40 Rev J Figure B 2 Models FL7218 FL7040 FL7060 ...
Page 54: ......